Design of a Novel Multifunction Decision Support Display for Anesthesia Care: AlertWatch® OR.
Tremper, Kevin K; Mace, Jenny J; Gombert, Jan M; Tremper, Theodore T; Adams, Justin F; Bagian, James P.
Afiliación
  • Tremper KK; Department of Anesthesiology, University of Michigan Medicine, 1500 E. Medical Center Drive, Ann Arbor, MI, 48109, USA. ktremper@umich.edu.
  • Mace JJ; Department of Anesthesiology, University of Michigan Medicine, 1500 E. Medical Center Drive, Ann Arbor, MI, 48109, USA.
  • Gombert JM; Department of Anesthesiology, University of Michigan Medicine, 1500 E. Medical Center Drive, Ann Arbor, MI, 48109, USA.
  • Tremper TT; Department of Anesthesiology, University of Michigan Medicine, 1500 E. Medical Center Drive, Ann Arbor, MI, 48109, USA.
  • Adams JF; AlertWatch Headquarters, 330 E. Liberty Street, Ann Arbor, MI, 48104, USA.
  • Bagian JP; Department of Anesthesiology, University of Michigan Medicine, 1500 E. Medical Center Drive, Ann Arbor, MI, 48109, USA.
BMC Anesthesiol ; 18(1): 16, 2018 02 05.
Article en En | MEDLINE | ID: mdl-29402220
ABSTRACT

BACKGROUND:

This paper describes the design of a multifunction alerting display for intraoperative anesthetic care. The design was inspired by the multifunction primary flight display used in modern aviation.

RESULTS:

The display retrieves live data from multiple sources; the physiologic monitors, the anesthesia information management system, the laboratory values and comorbidities from patient's problem summary list, medical history or history & physical. This information is integrated into a display composed of readily identifiable icons of organ systems, which are color coded to signify normal range, marginal range, abnormal range (by green, yellow, red respectively) and orange outlines for comorbidities/risk factors. There are dozens of text alerts, which can be presented as black text (informational), red text (important information) and red scrolling text (highest importance information). The alerts are derived from current standards in the literature and some involve complex calculations being conducted in the background.

CONCLUSIONS:

The goal of such a system is to improve the quality and safety of anesthetic care by providing enhanced situational awareness in a fashion analogous to the "glass cockpit" and its primary flight display which has improved aviation safety.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Monitoreo Intraoperatorio / Sistemas de Apoyo a Decisiones Clínicas / Diseño de Equipo / Anestesia Tipo de estudio: Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Año: 2018 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Monitoreo Intraoperatorio / Sistemas de Apoyo a Decisiones Clínicas / Diseño de Equipo / Anestesia Tipo de estudio: Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Año: 2018 Tipo del documento: Article