Your browser doesn't support javascript.
loading
Effect of each component of a LINAC therapy head on neutron and photon spectra.
Khabaz, Rahim.
Afiliación
  • Khabaz R; Physics Department, Faculty of Sciences, Golestan University, Postal code: 49138-15739, Gorgan, Iran. Electronic address: r.khabaz@gu.ac.ir.
Appl Radiat Isot ; 139: 40-45, 2018 Sep.
Article en En | MEDLINE | ID: mdl-29704704
ABSTRACT
Linear accelerators (LINACs) are widely applied in radiotherapy for their versatility and flexibility. Monte Carlo simulations were made to find the neutron and photon spectra at the isocenter (IC) of a LINAC operating at 10, 15, 18, and 24 MV by the MCNPX code. A detailed model of the LINAC head, consisting of flattening filter, secondary collimator, primary collimator, and multi-leaf collimator were used in the calculations. The effect of eliminating any of these components on contamination of a neutron spectrum and a photon spectrum was assessed. Photon and neutron ambient equivalent doses were found, and comparisons were made for the various structures. Lethargy neutron spectra at the IC were compared with spectra computed with the function reported by Tosi et al., which describes well neutron spectra for the energy region beyond 1 MeV, although tending to undervalue energy spectra below 1 MeV. The findings show that the photon and neutron fluences are enhanced when eliminating a LINAC component. The neutron and photon doses increased except when removing the primary collimator.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Aceleradores de Partículas / Radiocirugia / Fotones / Neutrones Tipo de estudio: Health_economic_evaluation / Prognostic_studies Límite: Humans Idioma: En Año: 2018 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Aceleradores de Partículas / Radiocirugia / Fotones / Neutrones Tipo de estudio: Health_economic_evaluation / Prognostic_studies Límite: Humans Idioma: En Año: 2018 Tipo del documento: Article