Reshaping bicuspid aortic valve stenosis with an hourglass-shaped balloon for transcatheter aortic valve replacement: A pilot study.
Zhao, Zhen-Gang; Feng, Yuan; Liao, Yan-Biao; Li, Yi-Jian; Xiong, Tian-Yuan; Ou, Yuan-Weixiang; Wang, Zi-Jie; Wei, Jia-Fu; Peng, Yong; Wei, Xin; Xu, Yuan-Ning; Pedersen, Wesley R; Chen, Mao; Leon, Martin B.
Afiliación
  • Zhao ZG; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Feng Y; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Liao YB; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Li YJ; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Xiong TY; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Ou YW; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Wang ZJ; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Wei JF; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Peng Y; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Wei X; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Xu YN; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Pedersen WR; Minneapolis Heart Institute Foundation, Abbott Northwestern Hospital, Twin Cities Heart Foundation, Minneapolis, Minnesota.
  • Chen M; Department of Cardiology, West China Hospital, Sichuan University, Chengdu, China.
  • Leon MB; Division of Cardiology, Columbia University Medical Center and the Cardiovascular Research Foundation, New York, New York.
Catheter Cardiovasc Interv ; 95 Suppl 1: 616-623, 2020 02.
Article en En | MEDLINE | ID: mdl-31943783
ABSTRACT

OBJECTIVES:

We evaluated the safety and usefulness of preparatory anatomical reshaping with a geometric hourglass-shaped balloon to optimize transcatheter aortic valve replacement (TAVR) outcomes in bicuspid aortic valve (BAV) stenosis.

BACKGROUND:

TAVR has been increasingly performed for BAV stenosis; however, technical challenges remain. Procedural results are suboptimal given unfavorable valvular anatomies.

METHODS:

Eligible patients with BAV stenosis were enrolled to undergo aortic valve predilatation with the hourglass-shaped TAV8 balloon before TAVR using the self-expandable Venus A-Valve. Procedural details and outcomes were compared to a sequential group of patients with BAV who underwent TAVR with the same device following preparatory dilatation using a cylindrical balloon.

RESULTS:

A total of 22 patients were enrolled in the TAV8 group and 53 were included in the control group. Valve downsizing was less common in the TAV8 group (36.4 vs. 67.9%; p = .012). Stable valve release and optimal implant depth were consistently achieved in the TAV8 group with no requirement for a second valve (0 vs. 17.0%; p = .039) and with higher device success rates (100.0 vs 77.4%; p = .014). Residual aortic regurgitation graded as ≥mild was less common in the TAV8 group (13.6 vs 45.3%; p = .009). Mortality was similar (0 vs. 3.8%; p = 1); no major/disabling stroke or conversion to open-heart surgery was seen in either group within 30 days.

CONCLUSIONS:

Compared with standard cylindrical balloon valvuloplasty, preparatory reshaping with the hourglass-shaped balloon before self-expandable TAVR in BAV was associated with significantly better procedural results and may encourage more promising outcomes.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Prótesis Valvulares Cardíacas / Valvuloplastia con Balón / Reemplazo de la Válvula Aórtica Transcatéter / Enfermedad de la Válvula Aórtica Bicúspide Tipo de estudio: Etiology_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Año: 2020 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Prótesis Valvulares Cardíacas / Valvuloplastia con Balón / Reemplazo de la Válvula Aórtica Transcatéter / Enfermedad de la Válvula Aórtica Bicúspide Tipo de estudio: Etiology_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Año: 2020 Tipo del documento: Article