Otto, Alexander; Siebenlist, Sebastian; Baldino, Joshua B; Murphy, Matthew; Muench, Lukas N; Mehl, Julian; Obopilwe, Elifho; Cote, Mark P; Imhoff, Andreas B; Mazzocca, Augustus D.
Afiliación
  • Otto A; Department of Orthopaedic Surgery, UConn Musculoskeletal Institute, University of Connecticut, Farmington, CT, USA.
  • Siebenlist S; Department of Orthopaedic Sports Medicine, Klinikum rechts der Isar, Technical University of Munich, Munich, Germany.
  • Baldino JB; Department of Trauma, Orthopaedic, Plastic and Hand Surgery, University Hospital of Augsburg, Augsburg, Germany.
  • Murphy M; Department of Orthopaedic Sports Medicine, Klinikum rechts der Isar, Technical University of Munich, Munich, Germany.
  • Muench LN; Department of Orthopaedic Surgery, UConn Musculoskeletal Institute, University of Connecticut, Farmington, CT, USA.
  • Mehl J; Department of Orthopaedic Surgery, UConn Musculoskeletal Institute, University of Connecticut, Farmington, CT, USA.
  • Obopilwe E; Department of Orthopaedic Sports Medicine, Klinikum rechts der Isar, Technical University of Munich, Munich, Germany.
  • Cote MP; Department of Orthopaedic Sports Medicine, Klinikum rechts der Isar, Technical University of Munich, Munich, Germany.
  • Imhoff AB; Department of Orthopaedic Surgery, UConn Musculoskeletal Institute, University of Connecticut, Farmington, CT, USA.
  • Mazzocca AD; Department of Orthopaedic Surgery, UConn Musculoskeletal Institute, University of Connecticut, Farmington, CT, USA.
JSES Int ; 4(4): 833-837, 2020 Dec.
Article en En | MEDLINE | ID: mdl-33345223
ABSTRACT

HYPOTHESIS:

The purpose of this study was to biomechanically evaluate onlay subpectoral long head of the biceps (LHB) tenodesis with all-suture anchors and unicortical buttons in cadaveric specimens.

METHODS:

After evaluation of bone mineral density, 18 fresh-frozen, unpaired human cadaveric shoulders were randomly assigned to 2 groups One group received an onlay subpectoral LHB tenodesis with 1 all-suture anchor, whereas the other group received a tenodesis with 1 unicortical button. The specimens were mounted in a servo-hydraulic material testing system. Tendons were initially loaded from 5 N to 100 N for 5000 cycles at 1 Hz. Displacement of the repair constructs was observed with optical tracking. After cyclic loading, each specimen was loaded to failure at a rate of 1 mm/s.

RESULTS:

The mean displacement after cyclic loading was 6.77 ± 3.15 mm in the all-suture anchor group and 8.41 ± 3.17 mm in the unicortical button group (P = not significant). The mean load to failure was 278.05 ± 38.77 N for all-suture anchor repairs and 291.36 ± 49.69 N for unicortical button repairs (P = not significant). The most common mode of failure in both groups was LHB tendon tearing. There were no significant differences between the 2 groups regarding specimen age (58.33 ± 4.37 years vs. 58.78 ± 5.33 years) and bone mineral density (0.50 ± 0.17 g/cm2 vs. 0.44 ± 0.19 g/cm2).

CONCLUSION:

All-suture anchors and unicortical buttons are biomechanically equivalent in displacement and load-to-failure testing for LHB tenodesis. All-suture anchors can be considered a validated alternative for onlay subpectoral LHB tenodesis.