Your browser doesn't support javascript.
loading
NEMA NU 2-2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT.
Zeimpekis, Konstantinos G; Kotasidis, Fotis A; Huellner, Martin; Nemirovsky, Alexandra; Kaufmann, Philipp A; Treyer, Valerie.
Afiliación
  • Zeimpekis KG; Department of Nuclear Medicine, University Hospital Zurich, University of Zurich, Rämistrasse 100, 8091, Zurich, Switzerland. konstantinos.zeimpekis@insel.ch.
  • Kotasidis FA; Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Freiburgstrasse 18, 3010, Bern, Switzerland. konstantinos.zeimpekis@insel.ch.
  • Huellner M; GE Healthcare, Waukesha, WI, USA.
  • Nemirovsky A; Department of Nuclear Medicine, University Hospital Zurich, University of Zurich, Rämistrasse 100, 8091, Zurich, Switzerland.
  • Kaufmann PA; GE Healthcare, Waukesha, WI, USA.
  • Treyer V; Department of Nuclear Medicine, University Hospital Zurich, University of Zurich, Rämistrasse 100, 8091, Zurich, Switzerland.
Eur J Nucl Med Mol Imaging ; 49(9): 3023-3032, 2022 07.
Article en En | MEDLINE | ID: mdl-35284970
ABSTRACT

PURPOSE:

The DMI PET/CT is a modular silicon photomultiplier-based scanner with an axial field-of-view (FOV) between 15 and 25 cm depending on ring configuration (3, 4, or 5 rings). A new generation of the system includes a reengineered detector module, featuring improved electronics and an additional 6th ring, extending the axial FOV to 30 cm. We report on the performance evaluation of the 6-ring upgraded Generation 2 (Gen2) system while values are also reported for the 5-ring configuration of the very same system prior to the upgrade.

METHODS:

PET performance was evaluated using the NEMA NU 2-2018 standard for spatial resolution, sensitivity, image quality, count rate performance, timing resolution, and image co-registration accuracy. Patient images were used to assess image quality.

RESULTS:

The average system sensitivity was measured at 32.76 cps/kBq (~ 47% increase to 5 rings at 22.29 cps/kBq) while noise equivalent count rate peaked at 434.3 kcps corresponding to 23.6 kBq/mL (~ 60% increase to Generation 1 (Gen1) and 39% to Gen2 5 rings). Contrast recovery ranged between 54.5 and 85.8% similar to 5 rings, while the 6 rings provided lower background variability (2.3-8.5% for 5 rings vs 1.9-6.8% for 6 rings) and lower lung error (4.0% for the 5 rings and 3.16% for the 6 rings). Transverse/axial full width at half-maximum (FWHM) at 1 cm (3.79/4.26 mm) and 10 cm (4.29/4.55 mm), scatter fraction (40.2%), energy resolution (9.63%), and time-of-flight (TOF) resolution (389.6 ps at 0 kBq/mL) were in line to previously reported values measured across different system configurations. Improved patient image quality is obtained with the 6 rings compared to the 5 rings, while image quality is retained even at reduced scan times, enabling WB dynamic acquisitions.

CONCLUSIONS:

The higher sensitivity of the 6-ring DMI compared to the 5-ring configuration may lead to improved image quality of clinical images at reduced scan time. Additionally, it could equally be used to allow improved temporal sampling and/or reduced overall scan time in dynamic acquisitions. Conversely, temporal sampling and scan time could be traded per application to further drive injected dose at lower levels.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tomografía de Emisión de Positrones / Tomografía Computarizada por Tomografía de Emisión de Positrones Límite: Humans Idioma: En Año: 2022 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tomografía de Emisión de Positrones / Tomografía Computarizada por Tomografía de Emisión de Positrones Límite: Humans Idioma: En Año: 2022 Tipo del documento: Article