A Novel Three-dimensional Planning Tool for Selective Clamping During Partial Nephrectomy: Validation of a Perfusion Zone Algorithm.
Eur Urol
; 83(5): 413-421, 2023 05.
Article
en En
| MEDLINE
| ID: mdl-36737298
ABSTRACT
BACKGROUND:
Selective clamping during robot-assisted partial nephrectomy (RAPN) requires extensive knowledge on patient-specific renal vasculature, obtained through imaging.OBJECTIVE:
To validate an in-house developed perfusion zone algorithm that provides patient-specific three-dimensional (3D) renal perfusion information. DESIGN, SETTING, ANDPARTICIPANTS:
Between October 2020 and June 2022, 25 patients undergoing RAPN at Ghent University Hospital were included. Three-dimensional models, based on preoperative computed tomography (CT) scans, showed the clamped artery's ischemic zone, as calculated by the algorithm. SURGICAL PROCEDURE All patients underwent selective clamping during RAPN. Indocyanine green (ICG) was administered to visualize the true ischemic zone perioperatively. Surgery was recorded for a postoperative analysis. MEASUREMENTS The true ischemic zone of the clamped artery was compared with the ischemic zone predicted by the algorithm through two metrics (1) total ischemic zone overlap and (2) tumor ischemic zone overlap. Six urologists assessed metric 1; metric 2 was assessed objectively by the authors. RESULTS ANDLIMITATIONS:
In 92% of the cases, the algorithm was sufficiently accurate to plan a selective clamping strategy. Metric 1 showed an average score of 4.28 out of 5. Metric 2 showed an average score of 4.14 out of 5. A first limitation is that ICG can be evaluated only at the kidney surface. A second limitation is that mainly patients with impaired renal function are expected to benefit from this technology, but contrast-enhanced CT is required at present.CONCLUSIONS:
The proposed new tool demonstrated high accuracy when planning selective clamping for RAPN. A follow-up prospective study is needed to determine the tool's clinical added value. PATIENTSUMMARY:
In partial nephrectomy, the surgeon has no information on which specific arterial branches perfuse the kidney tumor. We developed a surgeon support system that visualizes the perfusion zones of all arteries on a three-dimensional model and indicates the correct arteries to clamp. In this study, we validate this tool.Palabras clave
Texto completo:
1
Banco de datos:
MEDLINE
Asunto principal:
Procedimientos Quirúrgicos Robotizados
/
Neoplasias Renales
Tipo de estudio:
Observational_studies
/
Prognostic_studies
/
Risk_factors_studies
Límite:
Humans
Idioma:
En
Año:
2023
Tipo del documento: