Calibration method of crosshair light sharing PET detector with TOF and DOI capabilities.
Yoshida, Eiji; Obata, Fujino; Yamaya, Taiga.
Afiliación
  • Yoshida E; National Institutes for Quantum Science and Technology, 4-9-1 Anagawa, Inage-ku, Chiba 263-8555, Japan.
  • Obata F; National Institutes for Quantum Science and Technology, 4-9-1 Anagawa, Inage-ku, Chiba 263-8555, Japan.
  • Yamaya T; National Institutes for Quantum Science and Technology, 4-9-1 Anagawa, Inage-ku, Chiba 263-8555, Japan.
Biomed Phys Eng Express ; 9(5)2023 08 29.
Article en En | MEDLINE | ID: mdl-37586333
ABSTRACT
Objective. A crosshair light sharing (CLS) PET detector as a TOF-DOI PET detector with high spatial resolution has been developed. To extend that work, a detector calibration method was developed to achieve both higher coincidence resolving time (CRT) and DOI resolution.Approach. The CLS PET detector uses a three-layer reflective material in a two-dimensional crystal array to form a loop structure within a pair of crystals, enabling a CRT of about 300 ps and acquisition of DOI from multi-pixel photon counter (MPPC) output ratios. The crystals were 1.45 × 1.45 × 15 mm3fast LGSO, and the crystal array was optically coupled to an MPPC array. It is important to reduce as many inter-crystal scattering (ICS) events as possible in advance for the accurate detector calibration. DOI information is also expected to improve the CRT because it can estimate the time delay due to the detection depth of crystals.Main results. Using crystal identification and light collection rate of the highest MPPC output reduces the number of ICS events, and CRT is improved by 26%. In addition, CRT is further improved by 13% with a linear correction of time delay as a function of energy. The DOI is ideally estimated from the output ratio of only the MPPC pairs optically coupled to the interacted crystals, which is highly accurate, but the error is large due to light leakage in actual use. The previous method, which also utilizes light leakage to calculate the output ratio, is less accurate, but the error can be reduced. Using the average of the two methods, it is possible to improve the DOI resolution by 12% while maintaining the smaller error.Significance. By applying the developed calibration method, the CLS PET detector achieves the CRT of 251 ps and the DOI resolution of 3.3 mm.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Fotones / Tomografía de Emisión de Positrones Tipo de estudio: Prognostic_studies Idioma: En Año: 2023 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Fotones / Tomografía de Emisión de Positrones Tipo de estudio: Prognostic_studies Idioma: En Año: 2023 Tipo del documento: Article