The Laskin aerosol generator.
J Toxicol Environ Health ; 4(4): 661-70, 1978 Jul.
Article en En | MEDLINE | ID: mdl-682215
ABSTRACT
This paper describes construction details and operating characteristics of a nebulizer developed by Sidney Laskin and used over a period of 30 yr in various laboratories to generate respirable aerosols for whole-animal inhalation exposure studies. Under the proper operating conditions, the device is capable of producing nearly monodisperse aerosols in the respirable size range (1.5 micron volume median diameter with a geometric standard deviation of 1.1) for long periods of time