Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters

Database
Language
Affiliation country
Publication year range
1.
Diabetes Obes Metab ; 19(4): 553-561, 2017 04.
Article in English | MEDLINE | ID: mdl-27981804

ABSTRACT

AIMS: To evaluate the safety, efficacy and need for remote monitoring of the MD-Logic closed-loop system during short-term overnight use at home. METHODS: Seventy-five patients (38 male; aged 10-54 years; average A1c, 7.8% ± 0.7%, 61.8 ± 7.2 mmol/mol) were enrolled from 3 clinical sites. Patients were randomly assigned to participate in 2 overnight crossover periods, each including 4 consecutive nights, 1 under closed-loop control and 1 under sensor-augmented pump (SAP) therapy in the patient's home. Both study arms were supervised using a remote-monitoring system in a blinded manner. Primary endpoints were time spent with glucose levels below 70 mg/dL and percentage of nights in which mean overnight glucose levels were within 90 to 140 mg/dL. RESULTS: The median [interquartile range] percentage of time spent in hypoglycaemia was significantly lower on nights when MD-Logic was used, compared to SAP therapy (2.07 [0, 4.78] and 2.6 [0, 10.34], respectively; P = .004) and the percentage of individual nights with a mean overnight glucose level in target was significantly greater (75 [42, 75] and 50 [25,75], respectively; P = .008). The time spent in target range was increased by a median of 28% (P = .001), with the same amount of insulin (10.69 [7.28, 13.94] and 10.41[6.9, 14.07], respectively; P = .087). The remote monitoring triggered calls for hypoglycaemia at twice the rate during SAP therapy compared to closed-loop control (62 and 29, respectively; P = .002). CONCLUSIONS: The MD-Logic system demonstrated a safe and efficient profile during overnight use by children, adolescents and adults with type 1 diabetes and, therefore, provides an effective means of mitigating the risk of nocturnal hypoglycaemia.


Subject(s)
Blood Glucose Self-Monitoring/instrumentation , Diabetes Mellitus, Type 1/drug therapy , Hypoglycemic Agents/administration & dosage , Insulin Infusion Systems , Insulin/administration & dosage , Adolescent , Adult , Blood Glucose/analysis , Child , Cross-Over Studies , Diabetes Mellitus, Type 1/blood , Diabetes Mellitus, Type 1/complications , Drug Chronotherapy , Female , Humans , Hypoglycemia/blood , Hypoglycemia/drug therapy , Hypoglycemia/etiology , Male , Middle Aged , Single-Blind Method , Telemetry/methods , Time Factors , Treatment Outcome , Young Adult
SELECTION OF CITATIONS
SEARCH DETAIL