ABSTRACT

A fast liquid chromatographic method with tandem diode array-fluorescence detection for the simultaneous determination of in total 17 opium alkaloids and opioids is presented. Blank blood and urine samples (1 ml) were spiked with different concentrations of a standard mixture, as well as with the internal standard, butorphanol (2000 ng/ml). After solid-phase extraction, based on weak cation exchange (Bond Elut CBA SPE columns), the extracts were examined by HPLC-DAD-FL. By using a "high-speed" phenyl column (53 x 7.0 mm I.D., particle size 3 microm) eluted with a gradient system (A: water-methanol (90:10, v/v), B: methanol, both containing 25 mM triethylammoniumformate (pH(A) = 4.5)) all compounds could be baseline separated within 12 min. The method was validated and its applicability was demonstrated by the analysis of real-time forensic cases.


Subject(s)
Chromatography, High Pressure Liquid/methods , Narcotics/analysis , Opium/analysis , Spectrometry, Fluorescence/methods , Spectrophotometry, Ultraviolet/methods , Chromatography, Ion Exchange , Narcotics/blood , Narcotics/urine , Opium/blood , Opium/urine , Reference Standards , Reproducibility of Results , Sensitivity and Specificity