The conventional treatment of staphylococcal endocarditis requires in-patient administration, is inconvenient, and is potentially toxic. Increasing experience with well-absorbed, well-tolerated and highly active agents such as the new quinolones has prompted interest in their use as therapeutic alternatives for the treatment of such infections. We describe a case of staphylococcal endocarditis which failed to respond to conventional therapy, but where the addition of moxifloxacin, an 8-methoxyquinolone, was curative.
Subject(s)
Anti-Infective Agents/therapeutic use , Aza Compounds , Endocarditis, Bacterial/drug therapy , Fluoroquinolones , Quinolines , Staphylococcus aureus/isolation & purification , Adult , Endocarditis, Bacterial/microbiology , Humans , Male , Moxifloxacin , Staphylococcal Infections/drug therapyABSTRACT
A domestic, gaseous ozone generator was investigated for use in the decontamination of hospital side-rooms that have housed patients colonized with methicillin-resistant Staphylococcus aureus (MRSA). Three models of bacterial contamination were used. These were exposed to ozone generation in a standard hospital side-room for 4 and 7 h. A methicillin-sensitive and a methicillin-resistant strain of S. aureus were compared. Ozone concentrations of 0.14 ppm were reached, levels which are sufficient to cause mild pulmonary toxicity. Bacterial counts were reduced in the vicinity of the gas generator in most instances, but the effect elsewhere in the room was, at best, limited. MRSA appeared more resistant to the effects of ozone than methicillin-sensitive S. aureus. We conclude that the device tested would be inadequate for the decontamination of such hospital side-rooms.