Annu Int Conf IEEE Eng Med Biol Soc ; 2018: 490-493, 2018 Jul.
Article in English | MEDLINE | ID: mdl-30440441

ABSTRACT

Atrial tachycardia and atrial flutter are frequent arrhythmia that occur spontaneously and after ablation of atrial fibrillation. Depolarization waves that differ significantly from sinus rhythm propagate across the atria with high frequency (typically 140 to 220 beats per minute). A detailed and personalized analysis of the spread of depolarization is imperative for a successful ablation therapy. Thus, catheters with several electrodes are employed to measure multichannel electrograms inside the atria. Here we propose a new concept for spatio-temporal analysis of multichannel electrograms during atrial tachycardia and atrial flutter. It is based on the calculation of simultaneously active areas. The method allows to identify atrial tachycardia and to automatically distinguish between subtypes of focal activity, micro-reentry and macro-reentry.


Subject(s)
Atrial Fibrillation/diagnosis , Atrial Flutter/diagnosis , Electrophysiologic Techniques, Cardiac , Tachycardia, Supraventricular/diagnosis , Atrial Fibrillation/physiopathology , Atrial Flutter/physiopathology , Heart Atria/physiopathology , Humans , Spatio-Temporal Analysis , Tachycardia, Supraventricular/physiopathology