Locally fabricated phototherapy devices (LFPDs) are widely used in Nigeria for the treatment of neonatal jaundice. Ours was a cross-sectional observational study of all LFPDs in major hospitals in Jos between January and March 2015. We evaluated a total of 24 LFPDs. The irradiance at the level of the baby was in the range of 2-23.9 µW/cm2/nm. Fourteen devices had the recommended irradiance of ≥10 µW/cm2/nm and none had irradiance in the intensive range. Decreasing distance from the baby, presence of reflectors and increasing number of flourecent tubes significantly contributed to higher irradiance. A combination of six tubes, presence of reflectors and a distance of 10 cm from the baby produced a mean irradiance of 23.40 µW/cm2/nm. The irradiance of LFPDs varies widely and can be improved by simple modifications.