1.
Anal Bioanal Chem ; 375(1): 120-4, 2003 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-12520447

RESUMEN

A multi-analyte detection system using a unique antibody (Ab) biochip is described. The Ab-based biochip, also referred to as the protein biochip, uses a sensor array based on a complementary metal oxide silicon (CMOS) integrated circuit. The Ab-biochip has a sampling platform of four-by-four microarrays of antibodies deposited onto a Nylon membrane substrate. The micro-arrayed antibodies can be interrogated simultaneously or sequentially using the biochip sensing array detector with the use of a diffractive optical element illuminating each antibody spot individually. The usefulness of the Ab biochip is illustrated by the measurements of immunoglobulin G (IgG) used as the model analyte system. The detection limit for Cy5-labeled IgG molecules was 13 pg.


Asunto(s)
Anticuerpos Monoclonales/análisis , Técnicas Biosensibles/métodos , Análisis por Matrices de Proteínas/métodos , Animales , Complejo Antígeno-Anticuerpo/análisis , Técnicas Biosensibles/instrumentación , Calibración , Diseño de Equipo , Cabras , Inmunoglobulina G/análisis , Ratones , Análisis por Matrices de Proteínas/instrumentación , Conejos , Ovinos