1.
Radiat Prot Dosimetry ; 93(4): 325-9, 2001.
Artículo en Inglés | MEDLINE | ID: mdl-11548359

RESUMEN

Medical procedures denoted as interventional radiology require operation near an X ray beam, which brings high dose exposures to the operators' hands. For the effectual control of their extremity doses, a prototype of a real-time wrist dosemeter has been developed, hand dose monitor (HDM), based on a single silicon detector. Experiments were performed to test its response to diagnostic X rays. The HDM was highly sensitive and showed a linear response down to doses of a few tens of microsieverts. Though dose rate, energy and angular dependence of the response were observed in some extreme conditions, the HDM was proved to be of practical use if it was appropriately calibrated. Since an HDM enables personnel to check their hand doses on a real-time basis, it would enable medical staff to control the exposure themselves.


Asunto(s)
Mano/efectos de la radiación , Personal de Salud , Exposición Profesional , Radiología Intervencionista , Radiometría/instrumentación , Calibración , Humanos , Dosis de Radiación , Monitoreo de Radiación
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA