Colorectal Dis ; 6(2): 113-6, 2004 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-15008909

RESUMEN

OBJECTIVE: Dynamic graciloplasty is a treatment for intractable faecal incontinence. A gracilis muscle is transposed around the anus and stimulated with an implanted pulse generator (IPG). This in vitro study was designed to determine the safety of MRI in patients with implanted electrodes and pulse generators for dynamic graciloplasty. METHODS: Temperatures were measured with fiberoptic probes around the devices in a cadaver model. Current was measured with an oscilloscope connected to electrodes and IPG. Movement and IPG parameter setting were observed before, during and after testing. RESULTS: Minor temperature increase under 1 degrees C were observed around the electrodes. Amplitudes measured were within the range of -1 and +1 Volt. No movement or changing of IPG parameters was noted. CONCLUSION: Changes noted are well within physiological ranges. Dynamic graciloplasty is not a contraindication for high field strength MRI.


Asunto(s)
Terapia por Estimulación Eléctrica/instrumentación , Campos Electromagnéticos/efectos adversos , Imagen por Resonancia Magnética , Procedimientos Quirúrgicos Operativos/métodos , Cadáver , Electrodos Implantados , Estudios de Factibilidad , Incontinencia Fecal/cirugía , Calor/efectos adversos , Humanos , Músculo Esquelético/cirugía