Appl Radiat Isot ; 93: 76-81, 2014 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-24816175

RESUMEN

Method validation was performed to achieve the accreditation for our determination method of (210)Pb and (210)Po in water. A Pb(NO3)2 carrier is added to the sample and lead is precipitated with Na2SxH2O. (210)Po is co-precipitated and the extractive scintillation cocktail Polex(™) is used to determine (210)Po and (210)Pb. Uranium is also extracted by Polex(™). It can be removed by washing the precipitate with 1% HNO3. The ingrowth of (210)Pb from (222)Rn during transportation time must be calculated. It has to be subtracted from the original (210)Pb in the sample and taken into account for the calculation of the lower limit of detection.


Asunto(s)
Radioisótopos de Plomo/análisis , Polonio/análisis , Conteo por Cintilación/métodos , Contaminantes Radiactivos del Agua/análisis , Austria , Agua Subterránea/análisis , Humanos , Uranio/análisis , Abastecimiento de Agua/análisis , Abastecimiento de Agua/normas