Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 9 de 9
Filtrar
1.
Clin Res Cardiol ; 105(8): 657-665, 2016 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-26825328

RESUMO

PURPOSE: Electromagnetic interference (EMI) with implantable cardioverter defibrillators (ICDs) can cause oversensing and subsequently inappropriate ICD therapies. We retrospectively investigated the current incidence and clinical relevance of oversensing related EMI in a large cohort of ICD patients. METHODS: From January 2005 to April 2013, all ICD interrogations performed at our institution were analyzed for the occurrence of oversensing related EMI. EMI episodes were classified as clinically significant, potentially significant or of minor significance. To identify risk factors for EMI, we also analyzed different lead models in our cohort (integrated vs true bipolar leads). RESULTS: Data of 2940 ICD patients (mean age 63 ± 16 years, 2322 male patients, 7772 patient-years) were retrospectively analyzed for the occurrence of EMI. During the observation period, a total of 145 (hospital environment n = 97, non-hospital environment n = 48) episodes occurred and resulted in an overall EMI incidence, i.e. event rate, of 1.87 % per patient per year. Focusing on clinically significant or potentially significant episodes, the EMI incidence was 0.27 % per patient per year. Cox proportional hazards regression analysis did not reveal a statistically significant higher hazard of oversensing for patients with integrated bipolar leads compared to patients with true bipolar leads (HR = 2.21; 95 % CI 0.90-5.39; p = 0.083). CONCLUSIONS: Our data demonstrate that EMI continues to occur in everyday life. Patients should be well informed about the potential sources and risks of EMI but they need not be overly concerned since the risk of EMI-especially in a non-hospital environment-is low.


Assuntos
Arritmias Cardíacas/terapia , Desfibriladores Implantáveis , Cardioversão Elétrica/instrumentação , Campos Eletromagnéticos/efeitos adversos , Exposição Ambiental/efeitos adversos , Falha de Prótese , Idoso , Arritmias Cardíacas/diagnóstico , Arritmias Cardíacas/fisiopatologia , Técnicas Eletrofisiológicas Cardíacas , Feminino , Humanos , Masculino , Teste de Materiais , Pessoa de Meia-Idade , Modelos de Riscos Proporcionais , Desenho de Prótese , Estudos Retrospectivos , Medição de Risco , Fatores de Risco , Fatores de Tempo
3.
Europace ; 18(9): 1406-10, 2016 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-26589627

RESUMO

AIMS: This study presents and evaluates the impact of a new lowest-dose fluoroscopy protocol (Siemens AG), especially designed for electrophysiology (EP) procedures, on X-ray dose levels. METHODS AND RESULTS: From October 2014 to March 2015, 140 patients underwent an EP study on an Artis zee angiography system. The standard low-dose protocol was operated at 23 nGy (fluoroscopy) and at 120 nGy (cine-loop), the new lowest-dose protocol was operated at 8 nGy (fluoroscopy) and at 36 nGy (cine-loop). Procedural data, X-ray times, and doses were analysed in 100 complex left atrial and in 40 standard EP procedures. The resulting dose-area products were 877.9 ± 624.7 µGym² (n = 50 complex procedures, standard low dose), 199 ± 159.6 µGym² (n = 50 complex procedures, lowest dose), 387.7 ± 36.0 µGym² (n = 20 standard procedures, standard low dose), and 90.7 ± 62.3 µGym² (n = 20 standard procedures, lowest dose), P < 0.01. In the low-dose and lowest-dose groups, procedure times were 132.6 ± 35.7 vs. 126.7 ± 34.7 min (P = 0.40, complex procedures) and 72.3 ± 20.9 vs. 85.2 ± 44.1 min (P = 0.24, standard procedures), radiofrequency (RF) times were 53.8 ± 26.1 vs. 50.4 ± 29.4 min (P = 0.54, complex procedures) and 10.1 ± 9.9 vs. 12.2 ± 14.7 min (P = 0.60, standard procedures). One complication occurred in the standard low-dose and lowest-dose groups (P = 1.0). CONCLUSION: The new lowest-dose imaging protocol reduces X-ray dose levels by 77% compared with the currently available standard low-dose protocol. From an operator standpoint, lowest X-ray dose levels create a different, reduced image quality. The new image quality did not significantly affect procedure or RF times and did not result in higher complication rates. Regarding radiological protection, operating at lowest-dose settings should become standard in EP procedures.


Assuntos
Arritmias Cardíacas/cirurgia , Ablação por Cateter/métodos , Cineangiografia/métodos , Angiografia Coronária/métodos , Técnicas Eletrofisiológicas Cardíacas , Doses de Radiação , Exposição à Radiação/prevenção & controle , Radiografia Intervencionista/métodos , Adulto , Idoso , Arritmias Cardíacas/diagnóstico por imagem , Arritmias Cardíacas/fisiopatologia , Cateteres Cardíacos , Ablação por Cateter/efeitos adversos , Ablação por Cateter/instrumentação , Cineangiografia/efeitos adversos , Angiografia Coronária/efeitos adversos , Técnicas Eletrofisiológicas Cardíacas/efeitos adversos , Técnicas Eletrofisiológicas Cardíacas/instrumentação , Estudos de Viabilidade , Feminino , Fluoroscopia , Humanos , Masculino , Pessoa de Meia-Idade , Duração da Cirurgia , Exposição à Radiação/efeitos adversos , Radiografia Intervencionista/efeitos adversos , Estudos Retrospectivos , Fatores de Risco , Fatores de Tempo , Resultado do Tratamento
5.
J Cardiovasc Electrophysiol ; 26(1): 7-13, 2015 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-25091566

RESUMO

BACKGROUND: The role of dissociated pulmonary vein (DPV) activity after pulmonary vein isolation (PVI) is still poorly defined. We evaluated electrophysiological features and clinical impact on long-term outcome of DPV activity. METHODS: A total of 243 patients (mean age 63 ± 11 years; 63% males) undergoing PVI for paroxysmal atrial fibrillation (AF) were included. DPV activity was defined as a residual low frequency irregular PV rhythm. Patients were divided into Group 1 (presence of DPV activity; n = 65) or Group 2 (absence of DPV activity; n = 178). RESULTS: Of 936 isolated PVs, 112 PVs (12%) showed DPV activity. DPV activity was observed more frequently in PVs identified as AF triggers (P = 0.026). During follow-up (mean 12 ± 7 months), 15 of 65 patients of Group 1 (23%) and 57 of 178 patients of Group 2 (32%) had an arrhythmia recurrence (P = 0.23). At linear regression analysis, no independent predictor for clinical recurrence was identified. During the repeat ablation, 62 of 72 patients (86%) showed a recovered PV conduction without difference between the 2 groups. Clinically, all patients of Group 1 with PV reconnection (n = 13/15) had a recurrence of paroxysmal AF. In Group 2, 5 of 52 patients with reconnected PV developed non-PV related arrhythmias. CONCLUSION: DPV activity occurred in 12% of PVs after PVI and was observed more frequently in PVs identified as AF triggers. DPV activity was not predictive for AF recurrence during follow-up. PV-left atrium reconnection involving PVs with DPV activity leads to AF.


Assuntos
Fibrilação Atrial/cirurgia , Ablação por Cateter/efeitos adversos , Veias Pulmonares/cirurgia , Potenciais de Ação , Idoso , Fibrilação Atrial/diagnóstico , Fibrilação Atrial/fisiopatologia , Distribuição de Qui-Quadrado , Eletrocardiografia , Técnicas Eletrofisiológicas Cardíacas , Feminino , Frequência Cardíaca , Humanos , Estimativa de Kaplan-Meier , Modelos Lineares , Masculino , Pessoa de Meia-Idade , Modelos de Riscos Proporcionais , Veias Pulmonares/fisiopatologia , Recidiva , Fatores de Risco , Fatores de Tempo , Resultado do Tratamento
6.
J Cardiovasc Electrophysiol ; 24(4): 388-95, 2013 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-23252615

RESUMO

BACKGROUND: The endpoint of persistent atrial fibrillation (AF) ablation is still a matter of debate. The purpose of this study was to evaluate if sinus rhythm (SR) as endpoint of persistent AF ablation has a better long-term outcome compared to atrial tachycardia (AT) or AF at the end of the procedure. METHODS AND RESULTS: Between 2008 and 2011, 191 consecutive patients undergoing de novo catheter ablation for symptomatic persistent and long-standing persistent AF using a sequential ablation approach (including pulmonary vein isolation, ablation of complex fractionated electrograms and linear lesions) were included in the study. According to the result at the end of ablation procedure, patients were classified into 3 groups: patients with termination of AF into SR (Group 1, n = 62), patients with AT undergoing cardioversion (CV) (Group 2, n = 47), or patients with AF undergoing CV (Group 3, n = 82). The primary endpoint was freedom from any atrial tachyarrhythmia off antiarrhythmic drugs at 12 months. At 12 months, estimated proportions of patients free from any arrhythmia recurrence were 42% for Group 1, 13% for Group 2, and 25% for Group 3 (P = 0.002). In a Cox regression analysis only termination into SR was associated with a lower risk of arrhythmia recurrence (HR: 0.62; P = 0.04). CONCLUSION: If SR is achieved as endpoint of persistent and long-standing persistent AF ablation using a sequential ablation approach it is associated with the highest long-term single procedure success rate compared to AT or AF at the end of the procedure.


Assuntos
Fibrilação Atrial/cirurgia , Ablação por Cateter , Sistema de Condução Cardíaco/cirurgia , Veias Pulmonares/cirurgia , Adulto , Idoso , Antiarrítmicos/uso terapêutico , Fibrilação Atrial/diagnóstico , Fibrilação Atrial/fisiopatologia , Ablação por Cateter/efeitos adversos , Distribuição de Qui-Quadrado , Intervalo Livre de Doença , Cardioversão Elétrica , Eletrocardiografia Ambulatorial , Técnicas Eletrofisiológicas Cardíacas , Feminino , Sistema de Condução Cardíaco/fisiopatologia , Humanos , Masculino , Pessoa de Meia-Idade , Valor Preditivo dos Testes , Modelos de Riscos Proporcionais , Veias Pulmonares/fisiopatologia , Recidiva , Fatores de Risco , Taquicardia Supraventricular/etiologia , Taquicardia Supraventricular/fisiopatologia , Taquicardia Supraventricular/terapia , Fatores de Tempo , Resultado do Tratamento
7.
Am J Cardiol ; 101(3): 332-7, 2008 Feb 01.
Artigo em Inglês | MEDLINE | ID: mdl-18237595

RESUMO

Complex fractionated atrial electrographic (CFAE) catheter ablation is a new approach for the treatment of atrial fibrillation (AF). It is unclear if acute results of this approach correspond to long-term outcome. The purpose of this study was to prospectively assess acute and long-term successes of an ablation approach combining pulmonary vein isolation (PVI) and ablation of CFAE areas for treatment of persistent AF. PVI and ablation of CFAE areas were performed in 35 patients with persistent AF (30 men, 57+/-9 years of age). At the end of the ablation procedure AF had terminated in 23 of 35 patients (66%) by conversion to sinus rhythm (8 of 23 patients, 35%) or organization to atrial tachycardia (15 of 23 patients, 65%). AF persisted in 12 of 35 patients (34%). At the end of the follow-up period (19+/-12 months), sinus rhythm was present in 26 of 35 patients (74%), including 9 patients with a repeat procedure. This group of 26 patients consisted of 7 of 8 patients (88%) with acute sinus rhythm after the first ablation, 11 of 15 patients (73%) with organization, and 8 of 12 patients (66%) with ongoing AF (p=0.32). In conclusion, a combined approach of PVI and CFAE ablation in persistent AF leads to acute AF termination in 66% and long-term maintenance of sinus rhythm in 74% of cases. However, long-term outcome was not predictable by acute results of the ablation procedure.


Assuntos
Fibrilação Atrial/cirurgia , Ablação por Cateter/métodos , Técnicas Eletrofisiológicas Cardíacas , Veias Pulmonares/cirurgia , Idoso , Eletrocardiografia , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Estudos Prospectivos , Resultado do Tratamento
8.
J Cardiovasc Electrophysiol ; 18(10): 1039-46, 2007 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-17711435

RESUMO

INTRODUCTION: Complex fractionated atrial electrograms (CFAE) have been described as a new target for ablation of atrial fibrillation (AF). This prospective study evaluates the acute effects of CFAE ablation in patients with paroxysmal or persistent AF and analyzes the preferential anatomic sites where these effects occur. METHODS AND RESULTS: Ablation of CFAE was performed in 66 symptomatic patients (mean age of 58 +/- 12 years) with paroxysmal (n = 36) or persistent AF (n = 30). Termination or regularization of AF during ablation of CFAE was achieved in 56 of 66 patients (84%), with termination in 28 of 66 patients (42%) and regularization of AF in 28 of 66 patients (42%). Ablation of CFAE showed no effect in 10 of 66 patients (16%). Termination of AF occurred at 53 sites and AF regularization at 81 sites. The preferential sites of AF termination or regularization were found around the pulmonary veins (termination n = 15; regularization n = 22), at the anterior wall (termination n = 14; regularization n = 19) and at the interatrial septum (termination n = 8; regularization n = 17). CONCLUSION: Termination or regularization of AF was achieved acutely in 84% of patients by ablation of CFAE. The preferential sites of AF termination or regularization were found around the pulmonary veins, at the anterior wall of the LA and at the interatrial septum. These findings may have implications for future ablation concepts.


Assuntos
Fibrilação Atrial/diagnóstico , Fibrilação Atrial/terapia , Ablação por Cateter/métodos , Técnicas Eletrofisiológicas Cardíacas/métodos , Idoso , Fibrilação Atrial/fisiopatologia , Feminino , Seguimentos , Humanos , Masculino , Pessoa de Meia-Idade
SELEÇÃO DE REFERÊNCIAS