J Arthroplasty ; 30(3): 416-8, 2015 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-25453631

RESUMO

We report the early failure of five ceramic-on-ceramic hip resurfacings (CoCHRs). The ceramic used for the acetabular liner was a novel ceramic-composite (two thirds polyurethane and one third alumina ceramic). All cases were revised for increasing metal ion levels (blood cobalt 3.93-208.0 µg/l and chromium 1.57-17.5 µg/l) due to ceramic liner fracture and/or accelerated wear of the ceramic femoral head coating. Patients underwent bearing exchange and revision using primary hip arthroplasty implants at a mean of 3.0 years following CoCHR. Intraoperatively all patients had metallosis. At 1 to 2 years of follow-up blood metal ions normalized with no complications. We do not recommend this particular type of ceramic-on-ceramic bearing for hip resurfacing.


Assuntos
Artroplastia de Quadril/métodos , Prótese de Quadril/efeitos adversos , Falha de Prótese , Adulto , Óxido de Alumínio , Artroplastia de Quadril/instrumentação , Materiais Biocompatíveis , Cerâmica , Cromo/sangue , Cobalto/sangue , Corrosão , Feminino , Intoxicação por Metais Pesados , Humanos , Metais Pesados/sangue , Pessoa de Meia-Idade , Intoxicação/sangue , Intoxicação/etiologia