J Clin Anesth ; 37: 61-62, 2017 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-28235530

RESUMO

Bispectral index is an accepted depth of anaesthesia monitor for guiding intraoperative hypnotic agent administration. Frontalis EMG displayed on BIS monitor may increase due to twitching of frontalis muscle. EMG increases are also known to cause artefactual increases in BIS values. We report a case of artefactual increase of EMG and subsequently BIS values, due to electrical artefact from cranial nerve stimulator being used to identify the facial nerve. An explanation of the effect of stimulator signal on BIS EMG and BIS values has been provided.


Assuntos
Artefatos , Monitores de Consciência , Nervo Facial/fisiologia , Monitorização Intraoperatória/efeitos adversos , Estimulação Elétrica Nervosa Transcutânea , Anestesia Geral , Anestésicos Intravenosos/administração & dosagem , Craniotomia , Eletroencefalografia , Eletromiografia , Feminino , Fentanila/administração & dosagem , Humanos , Neoplasias Meníngeas/cirurgia , Meningioma/cirurgia , Pessoa de Meia-Idade , Monitorização Intraoperatória/instrumentação , Monitorização Intraoperatória/métodos , Fármacos Neuromusculares não Despolarizantes/administração & dosagem , Propofol/administração & dosagem , Brometo de Vecurônio/administração & dosagem