Acta Neurochir (Wien) ; 162(5): 1077-1079, 2020 05.
Artigo em Inglês | MEDLINE | ID: mdl-32034494

RESUMO

We report on a patient with thalamic deep brain stimulation (DBS) for essential tremor who was admitted to a stroke unit with transient vertigo, dysarthria, and gait disturbance. Transient ischemic attacks were assumed but fluctuating neurological symptoms persisted until presentation to a DBS center. Here, unstable high monopolar impedances of the right-hemispheric electrode contacts were detected. Surgical revision revealed a fracture of the pocket adaptor connecting this electrode to the impulse generator. Replacement resulted in stable impedances and remitted the transient neurological symptoms. Emergency and stroke doctors should be aware of neurological symptoms induced by technical dysfunctions in DBS.


Assuntos
Estimulação Encefálica Profunda/efeitos adversos , Erros de Diagnóstico , Eletrodos Implantados/efeitos adversos , Falha de Equipamento , Tremor Essencial/terapia , Ataque Isquêmico Transitório/diagnóstico , Idoso , Estimulação Encefálica Profunda/instrumentação , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Tálamo/fisiopatologia