[The determination of lorazepam in human urine by gas chromatography/nitrogen-phosphorus detector].
Jiang, Z L; Tan, J Y; Yao, L J; Xing, L M.
Afiliación
  • Jiang ZL; Dept. of Forensic Medicine, China Criminal Police College, Shenyang 110035, China.
Se Pu ; 19(4): 341-3, 2001 Jul.
Article en Zh | MEDLINE | ID: mdl-12545496
A method to assay lorazepam in human urine has been developed. After addition of hydroxyethylflurazepam (internal standard) and hydrolysis with beta-glucuronidase, the lorazepam and hydroxyethylflurazepam were extracted with ethyl ether at pH 10.8. The analysis was performed on an HP-5 capillary column with nitrogen-phosphorus detector(NPD). The detection limit and recovery of analytes in urine were 5 micrograms/L and (83.4 +/- 3.1)% respectively. The method was successfully applied to urine specimens collected from healthy human volunteers who have ingested 2 mg of lorazepam. The method was sensitive enough to assay urine specimen excreted at 32 h after taking the medicine by volunteers.
Asunto(s)
Buscar en Google
Bases de datos: MEDLINE Asunto principal: Ansiolíticos / Cromatografía de Gases / Lorazepam Idioma: Zh Revista: Se Pu Año: 2001 Tipo del documento: Article País de afiliación: China