Programming settings and recharge interval in a prospective study of a rechargeable sacral neuromodulation system for the treatment of overactive bladder.
Blok, Bertil; Van Kerrebroeck, Philip; de Wachter, Stefan; Ruffion, Alain; Van der Aa, Frank; Jairam, Ranjana; Perrouin-Verbe, Marie; Elneil, Sohier.
Afiliação
  • Blok B; Department of Urology, Erasmus M, edical Center, Rotterdam, The Netherlands.
  • Van Kerrebroeck P; Department of Urology, Maastricht University Medical Centre, Maastricht, The Netherlands.
  • de Wachter S; Department of Urology, University Hospital Antwerpen, Edegem, Belgium.
  • Ruffion A; Department of Urology, Hôpital Lyon Sud, Pierre Bénite, Lyon, France.
  • Van der Aa F; Department of Urology, UZ Leuven, Leuven, Belgium.
  • Jairam R; Department of Urology, Maastricht University Medical Centre, Maastricht, The Netherlands.
  • Perrouin-Verbe M; Department of Urology, University Hospital of Nantes, Nantes, France.
  • Elneil S; Department of Uro-Neurology, National Hospital of Neurology and Neurosurgery, London, United Kingdom.
Neurourol Urodyn ; 37(S2): S17-S22, 2018 02.
Article em En | MEDLINE | ID: mdl-29336058
ABSTRACT

AIMS:

The RELAX-OAB study is designed to confirm the safety, efficacy, and technical performance of the Axonics r-SNM System, a miniaturized, rechargeable SNM system approved in Europe and Canada for the treatment of bladder and bowel dysfunction. The purpose of this article is to describe study subjects' ability to charge the rechargeable neurostimulator and to document their neurostimulator program settings and recharge interval over time.

METHODS:

Fifty-one OAB patients were implanted in a single-stage procedure. These results represent the 3-month charging experience for 48 subjects who completed the 3-month follow-up. Recharge intervals were estimated using therapy stimulation settings and subject experience was evaluated using questionnaires.

RESULTS:

Forty-seven of forty-eight (98%) subjects were able to successfully charge their device prior to follow-up within 1-month post-implant. At 3-month post-implant, 98% of subjects were able to charge prior to their follow-up visit. Average stimulation amplitude across all subjects was 1.8 mA (±1.1 mA). A total of 69% of subjects had ≥14-day recharge intervals (time between charging) and 98% of subjects had ≥7-day recharge interval. No charging related adverse events occurred.

CONCLUSIONS:

Study subjects were able to charge the Axonics r-SNM System and stimulation settings provided 2 weeks of therapy between recharging for most subjects. Subject satisfaction indicates that subjects are satisfied with rechargeable SNM therapy.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Terapia por Estimulação Elétrica / Bexiga Urinária Hiperativa Tipo de estudo: Observational_studies País/Região como assunto: Europa Idioma: En Revista: Neurourol Urodyn Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Terapia por Estimulação Elétrica / Bexiga Urinária Hiperativa Tipo de estudo: Observational_studies País/Região como assunto: Europa Idioma: En Revista: Neurourol Urodyn Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Holanda