Intraoperative cell salvaging: ex vivo evaluation of two swab-washing methods.
Am J Vet Res
; 83(10)2022 Jul 30.
Article
em En
| MEDLINE
| ID: mdl-35895755
OBJECTIVE: To compare erythrocyte recovery by a cell salvage device between swab-washing by manual agitation or filtration. SAMPLE: 12 recently expired units of canine packed RBCs. PROCEDURE: The packed RBC units underwent quality analysis before donation from a pet blood bank. Each unit was volume-expanded with anticoagulant and subsequently divided into 2 equal aliquots used to soak surgical swabs before washing. Two different swab-washing techniques were evaluated-standard swab-washing-manual agitation (SW-MA) and swab-washing-filtration (SW-F)-with a novel prototype device. The resulting bloody fluid was processed using the Cell Saver Elite Autotransfusion System (Haemonetics). The volume, manual PCV, CBC, and RBC mass, calculated as the product of the volume and PCV, were measured before and after salvaging. Last, the RBC mass recovery was recorded as a percentage. RESULTS: The RBC mass recovered from SW-MA and SW-F averaged 85.73% and 83.99%, respectively. There was no significant difference in RBC recovery between the 2 methods (P = .52). CLINICAL RELEVANCE: SW-MA and SW-F recovered a similar quantity of RBCs from blood-soaked swabs in an ex vivo setting.
Texto completo:
1
Base de dados:
MEDLINE
Métodos Terapêuticos e Terapias MTCI:
Terapias_biologicas
/
Hemoterapia
Assunto principal:
Preservação de Sangue
/
Transfusão de Sangue Autóloga
Tipo de estudo:
Guideline
Idioma:
En
Revista:
Am J Vet Res
Ano de publicação:
2022
Tipo de documento:
Article
País de afiliação:
Reino Unido