Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros




Base de datos
Intervalo de año de publicación
1.
Neonatal Netw ; 33(3): 133-7, 2014.
Artículo en Inglés | MEDLINE | ID: mdl-24816873

RESUMEN

AIM: To study the effect of in-line filtering on in-line pressure measurement in a simulated infusion system. METHODS: The experimental setup consisted of a syringe pump, syringe, intravenous (IV) line, and extension line. For aqueous fluids, a 0.2-µm infusion filter was used, and for lipid emulsions, a 1.2-µm filter. Infusion speed varied from 0.5 to 10.0 mL/h. First, the effect of clamping the extension line was studied on reaching alarm threshold. Second, the effect of aqueous and lipid fluids was evaluated on in-line pressure measurements. RESULTS: In-line placement of an infusion filter did not prolong the time to alarm threshold after occlusion. During 24 hours of monitoring, lipid emulsions and all-in-one admixture only caused a moderate increase in in-line pressure. CONCLUSION: Placement of an infusion filter does not have a significant effect on the in-line pressure monitoring and has no adverse effect on detecting pressure-related complications of IV administration.


Asunto(s)
Alarmas Clínicas , Emulsiones Grasas Intravenosas , Filtración/instrumentación , Fluidoterapia/instrumentación , Fluidoterapia/enfermería , Bombas de Infusión , Unidades de Cuidado Intensivo Neonatal , Nutrición Parenteral Total/instrumentación , Nutrición Parenteral Total/enfermería , Transductores de Presión , Diseño de Equipo , Falla de Equipo , Humanos , Presión Hidrostática , Recién Nacido
SELECCIÓN DE REFERENCIAS