1.
J Endod ; 2024 Apr 18.
Artículo en Inglés | MEDLINE | ID: mdl-38642733

RESUMEN

INTRODUCTION: The purpose of this study was to evaluate the effect of side flattening of cutting flutes on the cyclic resistance and torsional resistance of nickel-titanium files. METHODS: Both novel flattened Platinum V.EU (PL) and standard non-flattened CC Premium V.EU (CC) rotaries were tested. For cyclic fatigue tests, all the files were rotated in an artificial root canal with a curvature of 45° and a radius of 6.06 mm at 300 rpm (n = 15 in each group). The number of cycles to failure (NCF) was calculated. For torsional tests, the files were rotated at 2 rpm clockwise until fracture occurred. The maximum torque value at fracture was measured and the toughness and distortion angle were computed. Subsequently, five fragments were randomly selected in each experiment, the cross-section and longitudinal direction of the fragments were photographed using a scanning electron microscope. An unpaired t-test was performed at a significance level of 95%. RESULTS: There was a statistically significant difference in NCF between CC and PL (P < .05). CC showed higher NCF than PL. There was no statistically significant difference between CC and PL with regards to the parameters related to torsional resistance (distortion angle, ultimate strength, and toughness) (P > .05). CONCLUSION: Within the limitations of this study, side flattening of the file did not improve cyclic resistance or torsional resistance of the files. As side flattening may reduce a file's cyclic resistance, such files should be used with caution in clinical practice.

2.
Aust Endod J ; 2024 Mar 25.
Artículo en Inglés | MEDLINE | ID: mdl-38528685

RESUMEN

The aim of this study was to assess the effect of ibuprofen sustained release (SR) oral premedication on the efficacy of buccal infiltration (BI) with intraoperative and postoperative pain after single-visit root canal treatment. Sixty patients diagnosed with symptomatic irreversible pulpitis and apical periodontitis in mandibular molar were divided into two groups. Group SR received ibuprofen SR 800 mg and group PL received placebo capsule 1 h before 3.6 mL articaine BI injection. Pain was recorded using a modified visual analogue scale and postoperatively at intervals 6, 24 and 48 h. Group SR showed a significantly higher anaesthetic success rate (73.3%) compared to group PL (46.7%) (p < 0.05). Intraoperative and postoperative pain was significantly higher in group PL compared to group SR (p < 0.05). Premedication of ibuprofen SR improved the efficacy of primary BI in mandibular molars with symptomatic irreversible pulpitis and decreased postoperative pain at 6 and 48 h.

3.
J Endod ; 50(4): 520-526, 2024 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-38311116

RESUMEN

INTRODUCTION: This study aimed to evaluate the effects from the memory-triple (MT) heat treatment on the fatigue resistance of the Reciproc by comparison with the file systems of same geometry. METHODS: Reciproc files subjected to MT heat treatment technology were designated as Group RMT and were compared with the original Reciproc (Group REC) and Reciproc Blue (Group REB). Each NiTi file from 3 groups (n = 15) was operated reciprocally with a repetitive up-and-down movement in the curved canal with 4 mm of pecking distance inside of the simulated canal at body temperature. When each file fractured, the time until fracture was recorded. The length of the fractured fragment was measured. Fractured fragments were observed under scanning electron microscope (SEM) to evaluate the topographic features of the surface. Differential scanning calorimetry (DSC) analysis was performed to estimate phase transformation temperatures. One-way analysis of variance and Duncan post hoc comparison were applied to compare among the groups at a significance level of 95%. RESULTS: RMT showed significantly higher fracture resistance (P < .05), whereas there was no difference in fatigue resistance between REC and REB. SEM examination showed the files from the 3 groups had similar topographic features. RMT showed a peak of austenite peak (Ap) at a temperature (52°C) higher than body temperature, whereas REC and REB showed Ap at 37 and 32°C, respectively. CONCLUSIONS: Under the condition of this study, the new heat treatment technique of MT technology could enhance the fatigue fracture resistance of the reciprocating files made of M-wire and Blue-wire.


Asunto(s)
Calor , Preparación del Conducto Radicular , Falla de Equipo , Estrés Mecánico , Instrumentos Dentales , Ensayo de Materiales , Diseño de Equipo , Titanio
4.
J Endod ; 50(2): 213-219, 2024 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-37924940

RESUMEN

INTRODUCTION: The aim of this study was to evaluate the torsional resistance, cyclic fatigue resistance, and bending stiffness of nickel-titanium (NiTi) file systems with different heat treatments and cross-sectional designs. METHODS: WaveOne Primary treated with memory-triple (MT) heat treatment (WOMT) was compared with WaveOne Primary (WO) and WaveOne Gold Primary (WOG). Torsional resistance test was performed using a customized device, and the distortion angle, ultimate strength, and toughness were evaluated. For cyclic fatigue resistance test, the instruments were reciprocated with continuous 4 mm up-and-down movement until fracture in a customized device, and the time to fracture was compared. Fracture surfaces of each group were examined under the scanning electron microscope. Bending stiffness was measured using a custom-made device. The results were analyzed using one-way analysis of variance and the Tukey's post hoc comparison at a significance level of 95%. RESULTS: WOMT showed higher ultimate strength and toughness than the other systems (P < .05). WOMT also showed highest cyclic fatigue resistance among the tested groups (P < .05). WO had the highest bending stiffness than others, whereas WOMT had a larger residual angle than others (P < .05). CONCLUSIONS: This new MT heat treatment technique makes NiTi file more flexible and improves its mechanical properties. In addition, the effect of heat treatment on flexibility was found to be more significant than that of the cross-sectional area.


Asunto(s)
Aleaciones , Calor , Titanio , Níquel , Instrumentos Dentales , Diseño de Equipo , Preparación del Conducto Radicular , Estrés Mecánico , Ensayo de Materiales , Torsión Mecánica
5.
J Endod ; 49(12): 1713-1721, 2023 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-37734586

RESUMEN

INTRODUCTION: This study aimed to evaluate the chemomechanical properties and biocompatibility of recently introduced premixed putty-type bioactive ceramic cements (PPBCs). METHODS: Including ProRoot MTA (PMTA) as a control, BC RRM fast-set putty (BCPT), Well-Root PT (WRPT), One-Fil PT (OFPT), and Endocem MTA premixed (ECPM) were compared to evaluate setting time, radiopacity, pH change, and microhardness. Biocompatibility on human dental pulp cells was compared using CCK-8 assay. Mineralization potential was evaluated using alkaline phosphatase activity, Alizarin Red S (ARS) staining, and quantitative real-time polymerase chain reaction with odontogenic gene marker. For data analysis, 1-way analysis of variance and Tukey's post hoc test were used at the significance level of 95%. RESULTS: Among the PPBCs, BCPT presented the longest (552 ± 27) setting time (minutes) and others showed significantly shorter time than PMTA (334 ± 22) (P < .05). WRPT (6.20 ± 0.54) and OFPT (5.82 ± 0.50) showed significantly higher radiopacity values (mmAl) and others showed similar value compared with PMTA (P > .05). All PPBCs showed high alkaline pH from fresh materials and tended to increase according to time periods from 30 minutes to 12 hours. ECPM showed the highest value of microhardness (81.62 ± 5.90), WRPT showed similar, and others showed lower than PMTA (P < .05). All PPBCs showed biocompatibility in CCK-8 assay. All PPBCs showed similar or better value compared with PMTA in ALP and ARS staining, and ALP and DSPP marker expression (P < .05). CONCLUSIONS: The PPBCs showed clinically acceptable chemomechanical properties and favorable mineralization potential.


Asunto(s)
Compuestos de Calcio , Materiales de Obturación del Conducto Radicular , Humanos , Compuestos de Calcio/farmacología , Compuestos de Calcio/química , Sincalida , Materiales de Obturación del Conducto Radicular/química , Odontogénesis , Cerámica , Silicatos/farmacología , Silicatos/química , Óxidos/química , Combinación de Medicamentos , Ensayo de Materiales , Compuestos de Aluminio/farmacología , Compuestos de Aluminio/química , Cementos Dentales/farmacología
6.
J Dent Sci ; 18(3): 1042-1046, 2023 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-37404604

RESUMEN

Background/purpose: Recently introduced calcium silicate-based bioceramic (CSBC) sealers require moisture for setting, thus this study aimed to compare the setting times of epoxy resin-and CSBC sealers under different test conditions. Materials and methods: Four CSBC sealers (CeraSeal, EndoSeal TCS, One-Fil, and Well-Root ST) were compared to an epoxy resin-based sealer (AH Plus). Each sealer was placed in a stainless-steel and gypsum molds on a glass slide. Sealer samples (n = 10 per group) were stored in an incubator at 95% humidity and 37 °C. A Gilmore needle with a total weight of 100 g and a 2.0-mm diameter were carefully placed vertically against the sealer, and the setting time was recorded when the needle no longer formed an indentation on the sealer surface. Statistical analysis was performed using two-way analysis of variance and Tukey parametric tests. The level of significance was set at 95%. Results: The setting time of all sealers in gypsum molds was significantly shorter than that in stainless-steel molds (P < 0.05). While AH Plus showed the longest setting time, EndoSeal TCS, One-Fil, and CeraSeal showed the shortest setting times when using gypsum molds among the five sealer types (P < 0.05). Conclusion: The results of this study indicate that CSBC sealers require moisture for setting; a lack of moisture results in a significant delay in setting time. Because the root canals contain moisture, it is necessary to experiment with the setting time of all types of sealers using gypsum molds to determine the biological condition of root canals.

7.
Aust Endod J ; 49(3): 675-683, 2023 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-37309727

RESUMEN

This article documents cases of single-visit pulp revascularisation for dens evaginatus and dens invaginatus, without using intracranial medicaments or antibiotics, aiming to provide a potentially applicable protocol for pulp revascularisation procedure in a single-visit. Two patients with chief complaints of pain and swelling visited a dental hospital. Radiographs revealed that the causative teeth had an open apex and periapical radiolucency, and the teeth were diagnosed as pulp necrosis and acute apical abscess or symptomatic apical periodontitis. For both cases, single-visit revascularisation was completed without intracanal medicaments or antibiotics. The patients were periodically recalled to evaluate periapical healing after treatment. The apical lesion healed, and the root dentin thickening was observed. The single-visit pulp revascularisation procedure without using specific intracanal medicaments can produce clinically favourable results for these dental anomalies.


Asunto(s)
Dens in Dente , Absceso Periapical , Humanos , Dens in Dente/terapia , Estudios de Seguimiento , Pulpa Dental , Absceso Periapical/terapia , Necrosis de la Pulpa Dental/terapia , Antibacterianos , Tratamiento del Conducto Radicular/métodos
8.
J Endod ; 49(7): 871-879, 2023 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-37225037

RESUMEN

INTRODUCTION: The aim of this study was to compare the physicochemical properties and biocompatibility of various calcium silicate-based bioceramic sealers (CSBSs). METHODS: Four recently developed CSBSs, including AH Plus Bioceramic Sealer (AHB), EndoSequence BC Sealer (ESB), TotalFill BC Sealer (TTB), and Bio-C Sealer (BIC), were compared with the epoxy resin-based sealer AH Plus (AHP). Their physical properties, including flow, setting time, radiopacity, dimensional stability, and pH, were evaluated according to the International Organization for Standardization (ISO) 6876. Their cytotoxicity in human periodontal ligament fibroblast (hPDLF) was assessed through the 3-(4,5-dimethylthiazol-2-yl)-2,5-diphenyltetrazolium bromide assay and compared. Additionally, cell attachment to the sealer surface was analyzed using green fluorescent protein and confocal laser scanning microscopy to evaluate cell viability. Data were analyzed using a one-way analysis of variance to determine the difference between groups for categorical variables, followed by Tukey's post hoc test at the significance level of 95%. RESULTS: The flow, setting time, and radiopacity of all tested CSBSs satisfied the ISO 6876/2012 standards. Further, these CSBSs showed shrinkage after immersion in distilled water for 30 days and complied with the ISO 6876/2001 requirements. The pH values of AHB, ESB, TTB, and BIC were greater than 11, whereas AHP had a pH of 6.69 after 4 weeks. CSBS showed excellent biocompatibility compared with that of AHP (P < .05). Confocal laser scanning microscopy showed that alive hPDLFs were attached well to all the tested CSBSs but not to AHP. CONCLUSIONS: CSBSs have similar physical characteristics within the ISO standards and higher biocompatibility than epoxy resin-based sealers.


Asunto(s)
Resinas Epoxi , Materiales de Obturación del Conducto Radicular , Humanos , Resinas Epoxi/farmacología , Resinas Epoxi/química , Materiales de Obturación del Conducto Radicular/farmacología , Materiales de Obturación del Conducto Radicular/química , Cavidad Pulpar , Ensayo de Materiales , Compuestos de Calcio/farmacología , Compuestos de Calcio/química , Silicatos/farmacología , Silicatos/química
9.
J Endod ; 49(1): 77-82, 2023 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-36375648

RESUMEN

INTRODUCTION: This study evaluated the effect of periodic changes in rotation speed on torsional stress and screw-in force using the dedicated alternative rotation technique (ART) motion of the EQ-M (Metabiomed, Cheongju, Korea) endodontic motor. METHODS: Two ART modes of the EQ-M motor in 2 alternative techniques (ART30 and ART50) and continuous rotation were compared using ProTaper Next X2 (Dentsply Sirona, Charlotte, NC) files and simulated resin blocks (n = 12 per group). ART30 and ART50 were operated by continuous rotation of 350 rpm for 360° and then rotated at 30% increased speed from the base speed for 180° and at 50% increased speed for 180°, respectively. Before the test, the simulated resin blocks were pre-enlarged using ProTaper Gold S1 and S2 (Dentsply Sirona) and fixed on a metal stage connected to the force- and torque-measuring unit. During shaping the simulated canal in an automatic up-and-down manner, the parameters of maximum torque, sum of torque, maximum screw-in force, and maximum apical force were measured. The data were statistically analyzed using 1-way analysis of variance and the Tukey post hoc comparison test at the 95% significance level. RESULTS: The ART30 and ART50 groups showed a lower maximum torque, sum of torques, screw-in force, and apical drive force than the continuous rotation group. There was no significant difference between the ART30 and ART50 groups (P > .05). When the screw-in force increased suddenly, the torque correspondently increased. CONCLUSIONS: Under the limitations of this study, the ART mode could reduce the torsional stress and apical forces of the screw-in during instrumentation in comparison with continuous rotation.


Asunto(s)
Aleaciones Dentales , Preparación del Conducto Radicular , Titanio , Diseño de Equipo , Torque , Tornillos Óseos , Estrés Mecánico , Ensayo de Materiales
10.
Materials (Basel) ; 15(15)2022 Jul 25.
Artículo en Inglés | MEDLINE | ID: mdl-35897577

RESUMEN

The purpose of this randomized controlled clinical trial was to assess the short-term clinical effectiveness of sealer-based root canal obturation using calcium-silicate-based sealer (CSBS) and epoxy-resin-based sealer (ERBS). A total of eighty patients (eighty teeth) were enrolled and seventy-two patients (seventy-two teeth) were randomly assigned into four different sealer groups: AH Plus (AH, Dentsply Sirona), ADseal (AD, Meta Biomed), CeraSeal (CS, Meta Biomed), or EndoSeal TCS (ES, Maruchi). At the first visit, root canal treatment was performed using ProTaper. Next, the NiTi file system and passive ultrasonic irritation was applied with 2.5% NaOCl. At the second visit, the teeth were obturated with a sealer-based obturation technique using the assigned sealer. The quality of filling obturation was evaluated with postoperative radiographs in terms of the presence of voids and sealer extrusion. The patients were recalled at 1-week, 1-month, and 3-month intervals, and pain levels were measured using visual analog scales (VAS), and clinical examination performed with percussion and palpation. Radiographs were also taken. The results were statistically compared by using the χ2 test or Fisher exact test. Among the 80 teeth included, 72 teeth received allocated intervention and 71 teeth were included in the analysis (98.6% recall). There was no significant difference among the sealers in void and sealer extrusion (p > 0.05). Postoperative pain was not indicated prominently regardless of the type of sealer. For a 3-month follow-up period, all cases were maintained successfully except for one case from AD. According to the findings, there was no significant difference between CSBS and ERBS in terms of postoperative pain or the healing process, and the variation in filling quality appears to be linked to the properties of each product.

11.
Materials (Basel) ; 15(2)2022 Jan 17.
Artículo en Inglés | MEDLINE | ID: mdl-35057383

RESUMEN

As the use of nickel-titanium (NiTi) file systems for root canal therapy has become popular; hence, knowledge and understanding of the characteristics of NiTi files is essential for dentists. Unintended sudden fracture can occur during root canal shaping, and it is important to understand the conditions that may cause instrument fractures. Torque is defined as the force required to rotate the NiTi file and can be considered of as a parameter for the stress generated. The endodontic engine maintains a constant rotational speed by adjusting torque regardless of the root canal conditions. The process of root canal shaping by rotary instruments is a series of actions that requires torque and generates stress to both the teeth and the NiTi instruments. The generated stress may induce the strain accumulation on NiTi instrument and the canal wall and lead to the development of microcrack in the instrument and dentinal wall. Therefore, understanding of torque and stress generated is important to prevent the fractures to the instrument and the teeth. This stress has been measured using various experimental approaches, including microcrack observation by using a microscope or computed tomography, attaching strain gauges to the teeth, and finite element analysis. This review focuses on the stress generated to the teeth and the instrument during instrumentation under various experimental conditions. The factors related to torque generation are also discussed.

12.
Aust Endod J ; 48(1): 131-137, 2022 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-34288256

RESUMEN

This study aimed to evaluate the effect of glide-path preparation and coronal pre-flaring on torque generation of rotary nickel-titanium files. Sixty mesiobuccal root canals of 3D printed resin teeth (Dentalike) were randomly divided into 4 groups (n = 15) according to the preparation method; Group 1: OneCurve only; Group 2: OneG and OneCurve; Group 3: OneFlare and OneCurve; Group 4: OneFlare, OneG, and OneCurve. During the final canal preparation procedure, the torque generated was recorded. The sum of generated torque and maximum torque were calculated. Statistical analysis was performed with one-way ANOVA and Duncan's post hoc comparison. The summed torque was reduced by glide-path preparation and coronal pre-flaring (P < 0.05), whereas the maximum torque was not reduced by the glide-path preparation (P > 0.05). It will be advantageous to create a glide-path with coronal pre-flaring to reduce the stress and reactive forces to the NiTi file and root dentin.


Asunto(s)
Cavidad Pulpar , Preparación del Conducto Radicular , Diseño de Equipo , Preparación del Conducto Radicular/métodos , Titanio , Raíz del Diente , Torque
13.
J Dent ; 117: 103939, 2022 02.
Artículo en Inglés | MEDLINE | ID: mdl-34953972

RESUMEN

OBJECTIVES: The present study examined the defect characteristics of clinically-used, discarded nickel titanium rotary retreatment instruments and analysed the impact of clinical use on their metallurgical properties. METHODS: 92 XP-endo Shaper (XPS; FKG Dentaire) and 20 XP-endo Finisher (XPFR; FKG Dentaire) instruments with structural deformation or separation were collected after retreatment from four endodontic clinics over a 20-month period. The types of defects and their relative locations were recorded. The lateral and fractured surfaces of the separated instruments were examined with scanning electron microscopy. Differential scanning calorimetry was used to investigate the thermal behaviour of new, deformed and fractured instruments. RESULTS: 77 (84%) XPS and 4 (20%) XPFR had an area with structural change while 15 (16%) XPS and 16 (80%) XPFR were fractured. All unfractured, deformed XPFR showed unwinding close to the coronal end of the flute. Fractures in XPS and XPFR were often close to the coronal end of the flutes or the expanding segments of the insruments. Most of the XPS fractures were torsional failure (67%) while XPFR failed predominantly by cyclic fatigue (81%). The austenite-finishing temperature of XPFR (40 °C) was higher than that of XPS (35 °C). Both XPS and XPFR exhibited 2-stage phase transformations. CONCLUSIONS: Torsional failure was more prevalent in XPS instruments and fatigue failure was more prevalent in XPFR instruments. amongst the investigated instruments, XPFR were more likely to separate without warning whereas XPS frequently exhibited plastic deformation. The latter may be used as a pre-separation forewarning sign during clinical retreatment. CLINICAL SIGNIFICANCE: The failure mode of XP-endo Shaper and XP-endo Finisher used clinically for retreatment appeared to be different. Plastic deformation, the forewarning sign of instrument seaparation, occurs when XP-endo instruments are used for retreatment.


Asunto(s)
Níquel , Preparación del Conducto Radicular , Aleaciones , Instrumentos Dentales , Diseño de Equipo , Níquel/química , Retratamiento , Titanio/química
14.
Materials (Basel) ; 14(18)2021 Sep 14.
Artículo en Inglés | MEDLINE | ID: mdl-34576520

RESUMEN

This study compared the torsional resistance of heat-treated nickel-titanium (NiTi) instruments under different temperature conditions. Four thermomechanically treated single-use NiTi rotary instruments were selected for this study: OneShape (OS), OneCurve (OC), WaveOne Gold (WOG) and HyFlex EDM (HFE). Each instrument was further subdivided by temperature into 2 subgroups. Maximum torque and the distortion angle until fracture occurred were evaluated. Differential scanning calorimetry analysis was performed to measure the phase transformation temperature. Statistical analysis was performed using a two-way ANOVA and t-test (p < 0.05). Fractured fragments were observed using scanning electron microscopy (SEM). The two-way ANOVA showed no significant differences for different temperature conditions. At both room (RT) and body temperature (BT), OS was predominantly austenite while HFE was martensite. OC and WOG were predominantly martensite at RT and mixed phase at BT. At BT, more than half of WOG was martensite, while half of OC was austenite. SEM examination showed no topographical differences between instruments in different temperature groups. In relation to a limitation of this study, there was no difference in torsional resistance of NiTi rotary instruments between the BT and RT conditions. This implies that clinicians do not need to consider a decrease of torsional resistance of heat-treated NiTi instruments at BT.

15.
J Endod ; 47(9): 1501-1506, 2021 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-34237386

RESUMEN

INTRODUCTION: This study aimed to compare the torque generated by 4 different files in root canals with 4 different curvature angles. METHODS: Four brands of nickel-titanium (NiTi) endodontic files were selected: WaveOne Primary (Dentsply Sirona, Ballaigues, Switzerland), WaveOne Gold Primary (Dentsply Sirona), ProTaper Universal F2 (Dentsply Sirona), and ProTaper Next X2 (Dentsply Sirona). A tempered steel block containing artificial canals with 4 different canal curvatures (15°, 25°, 35°, and 45°) was constructed. Each file was used according to the manufacturer's instructions in the dynamic model, with an added 15 axial up-and-down movements of 4 mm at the end of the canal. The generated torque was recorded, and the total and maximum torque values were measured. Two-way analysis of variance and the Duncan post hoc comparison test were performed at a significance level of 95%. RESULTS: A significant correlation between the curvature angle and the type of file system was observed (P < .05). As the degree of canal curvature increased, the generated total and maximum torque increased. At 15° and 25°, the NiTi files with reciprocating motion generated a higher total and maximum torque than files with continuous rotation. ProTaper Universal of conventional NiTi alloy showed the steepest increase in the generated total and maximum torque with the increasing curvature angle. The ProTaper Next file had the lowest torque values at the higher canal angles of 35° and 45° (P < .05). CONCLUSIONS: Despite the study limitations, it can be concluded that root canal curvature, design, and heat treatment of NiTi files and file kinematics affect the generated torque during instrumentation.


Asunto(s)
Níquel , Preparación del Conducto Radicular , Aleaciones , Aleaciones Dentales , Cavidad Pulpar , Diseño de Equipo , Rotación , Titanio , Torque
16.
Aust Endod J ; 47(3): 512-519, 2021 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-33894082

RESUMEN

The purpose of this study was to compare the physical properties and cytocompatibility of contemporary calcium silicate-based sealers. Four calcium silicate-based sealers (BrightEndo MTA sealer, CeraSeal, EndoSeal TCS and One-Fil) were compared to an epoxy resin-based sealer (AH Plus). Flow, setting time, radiopacity and dimensional change were evaluated according to ISO 6876 standards. Cytotoxicity on human periodontal ligament fibroblast (hPDLF) cells was compared for biological properties using MTT assay. The surface of the sealer was analysed using scanning electron microscopy to evaluate cell attachment. Flow and radiopacity of all sealers met ISO standards, while setting time and dimensional stability did not meet the ISO standards. Calcium silicate-based sealers showed favourable cytocompatibility, and hPDLF cells were well attached to the calcium silicate-based sealers. Calcium silicate-based sealers have clinically acceptable flow and radiopacity, and cytocompatibility. However, these sealers had longer setting time and higher dimensional change than those required by ISO 6876.


Asunto(s)
Proyectos de Investigación , Compuestos de Calcio , Humanos , Silicatos
17.
J Endod ; 47(2): 303-308, 2021 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-33091452

RESUMEN

INTRODUCTION: The purpose of this study was to investigate if machined springs in nickel-titanium (NiTi) rotary instruments can improve their mechanical properties. The bending and torsion properties were assessed using finite element (FE) model analysis. METHODS: A basic 3-dimensional file model without a spring was created with apical size #25, 25-mm full length, and 16-mm cutting flutes. Three other models were created with a spring machined into their shaft portion with different numbers of spring coils: standard (STspr), 10% more (INspr), and 10% less (DEspr). To compare the mechanical responses among the 4 FE models, file bending and torsion were simulated using FE analysis. RESULTS: Spring machined NiTi rotary instruments showed higher torsional resistance and less bending stiffness than the same instrument without. The spring machined models required more torque to bend or rotate the DEspr model than was required for the STspr and INspr models; however, the STSpr and INSpr models were similar. CONCLUSIONS: Within the limitations of this study, the FE analysis indicated that machining a spring into the shaft of NiTi rotary instruments improved torsional resistance and bending flexibility. Therefore, spring machining has the potential to increase the durability of the NiTi rotary instruments.


Asunto(s)
Níquel , Titanio , Aleaciones Dentales , Diseño de Equipo , Análisis de Elementos Finitos , Ensayo de Materiales , Preparación del Conducto Radicular , Estrés Mecánico , Torsión Mecánica
18.
J Endod ; 47(1): 88-93, 2021 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-32961213

RESUMEN

INTRODUCTION: This study aimed to compare the torsional resistances and fracture modes of WaveOne Gold (Dentsply Sirona, Ballaigues, Switzerland) and Reciproc Blue (VDW, Munich, Germany) using the repetitive torsional loading (RTL) method and the conventional single-rotation (STL) method. METHODS: A 3-mm file tip was fixed with a brass plate, and a torsional load was applied using a custom device. In the RTL method, the file was driven counterclockwise at 50 rpm until it achieved the preset torque of 0.5 Ncm; thereafter, it was returned to its original position. This recovery of the file to its original position was defined as 1 torsional loading cycle; the number of repetitive load cycles until fracture was counted. In the STL method, the files were rotated at a constant rate of 2 rpm in a counterclockwise direction until file fracture. The fragments fractured by the 2 methods were compared under a scanning electron microscope to examine the topographic features of the fractured surfaces and longitudinal aspects. RESULTS: With the RTL method, Reciproc Blue showed a higher number of repetitive load cycles until fracture than WaveOne Gold (P < .05). With the STL method, Reciproc Blue also had a higher ultimate strength than WaveOne Gold (P < .05). Scanning electron microscopic findings of the fractured specimens from the 2 test methods showed different features of torsional failure. CONCLUSIONS: Within the study limitations, both the RTL and STL methods conferred similar torsional resistance. Therefore, the clinically relevant RTL method with repetitive and reciprocation motion can be used for testing torsional resistance.


Asunto(s)
Oro , Preparación del Conducto Radicular , Instrumentos Dentales , Diseño de Equipo , Alemania , Ensayo de Materiales , Rotación , Titanio , Torque
19.
Materials (Basel) ; 13(22)2020 Nov 20.
Artículo en Inglés | MEDLINE | ID: mdl-33233623

RESUMEN

Nickel-titanium (NiTi) endodontic rotary instruments are used extensively in root canal procedures by both general dentists and specialists. However, their vulnerability to fracture is the major reason for clinicians' concern regarding their use. The objective of this study was to investigate the potential effects of spring machining of the file shaft on the fatigue and torsional resistances of NiTi rotary instruments. Three types of NiTi rotary systems with (S) and without (NS) spring machining were used in this study (n = 15 each): a spring file (SPR; #25/.06, SPR-S, SPR-NS), a ProTaper Next X2 (PTN; #25/variable taper, PTN-S, PTN-NS), and a ProTaper Gold F2 (PTG; #25/variable taper, PTG-S, PTG-NS). Spring machining was adjusted on the 6 mm of each file system's shaft via a laser cutting process. The number of rotation cycles until fracture (i.e., cyclic fatigue resistance), ultimate torsional strength, the distortion angle, and the toughness of each subgroup were estimated with specially designed devices. The results were analyzed using a paired t-test at a significance level of 95%. NiTi rotary instruments with spring machining exhibited a higher cyclic fatigue resistance than instruments without spring machining. The groups with spring machining exhibited a higher toughness and larger distortion angle than the groups without it (p < 0.05). In conclusion, spring machining on the shank of NiTi instruments may provide a stress-bearing area and attenuate the torsional and cyclic fatigue of NiTi rotary instruments.

20.
J Endod ; 46(8): 1125-1129, 2020 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-32497655

RESUMEN

INTRODUCTION: The aim of this study was to compare the buckling resistance of nickel-titanium (NiTi) instruments for glide path preparation depending on the test mode (static vs dynamic). METHODS: The conventional PathFile (PF; Dentsply Sirona, Ballaigues, Switzerland) and heat-treated ProGlider (PG, Dentsply Sirona) and WaveOne Gold Glider (WG, Dentsply Sirona) were evaluated. The instrument tips were placed in a small dimple prepared in an aluminum cube in a customized device. The file was then pushed in the axial direction at a 1.0-mm/s crosshead speed with rotation (dynamic mode) or without rotation (static mode). The dynamic mode of WG used its dedicated reciprocating rotation, whereas the others were rotated continuously at 300 rpm. The axial load and lateral buckling displacement were simultaneously measured. Data were analyzed statistically using 2-way analysis of variance (P = .05). RESULTS: The buckling resistance in the dynamic mode was higher than in the static mode for PG and WG (P < .05), whereas PF was not influenced by test modes (P > .05). In the dynamic mode, the PG required the highest buckling load followed by PF and WG (P < .05). In the static mode, the WG showed the lowest load (P < .05). The dynamic mode showed significantly more upper level lateral buckling displacement than in the static mode (P < .05). CONCLUSIONS: When the glide path preparation instruments moved in the dynamic mode as in clinical situations, the buckling resistance of the heat-treated NiTi glide path instruments was higher than in the static condition. The heat-treated instruments may have better buckling resistance than the conventional NiTi instrument in clinical situations.


Asunto(s)
Níquel , Titanio , Aleaciones Dentales , Diseño de Equipo , Ensayo de Materiales , Preparación del Conducto Radicular , Rotación