J Vet Cardiol ; 52: 68-71, 2024 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-38432142

RESUMEN

A seven-year-old male castrated Labrador Retriever presented emergently due to concern for pacemaker malfunction five years after successful transvenous pacemaker implantation to treat partial atrial standstill. On presentation, the dog's pulse rate was 30-50 beats per minute. An electrocardiogram showed no spontaneous atrial activity or paced ventricular activity. Pacemaker interrogation revealed an increased impedance of 7557 ohms, indicating a lead malfunction. Thoracic radiographs confirmed the lead was fractured and had excessive coiling. The transvenous pacing system was turned off, left in place, and an epicardial pacing system was implanted the following day. The dog was discharged with no perioperative complications. The dog eventually required escalated medical therapy for progressive cardiac disease and was euthanized two years after implantation of the replacement pacemaker. This manuscript illustrates a complete lead fracture and excessive lead coiling, which has not previously been detailed in veterinary medicine.


Asunto(s)
Enfermedades de los Perros , Falla de Equipo , Marcapaso Artificial , Perros/lesiones , Animales , Masculino , Marcapaso Artificial/veterinaria , Marcapaso Artificial/efectos adversos , Enfermedades de los Perros/terapia , Enfermedades de los Perros/diagnóstico por imagen , Falla de Equipo/veterinaria , Electrocardiografía/veterinaria