RESUMEN
PURPOSE: To investigate the effectiveness of an EPID-based 3D transit dosimetry system in detecting deliberately introduced errors during VMAT delivery. METHODS: An Alderson phantom was irradiated using four VMAT treatment plans (one prostate, two head-and-neck and one lung case) in which delivery, thickness and setup errors were introduced. EPID measurements were performed to reconstruct 3D dose distributions of "error" plans, which were compared with "no-error" plans using the mean gamma (γmean), near-maximum gamma (γ1%) and the difference in isocenter dose (ΔDisoc) as metrics. RESULTS: Out of a total of 42 serious errors, the number of errors detected was 33 (79%), and 27 out of 30 (90%) if setup errors are not included. The system was able to pick up errors of 5â¯mm movement of a leaf bank, a wrong collimator rotation angle and a wrong photon beam energy. A change in phantom thickness of 1â¯cm was detected for all cases, while only for the head-and-neck plans a 2â¯cm horizontal and vertical shift of the phantom were alerted. A single leaf error of 5â¯mm could be detected for the lung plan only. CONCLUSION: Although performed for a limited number of cases and error types, this study shows that EPID-based 3D transit dosimetry is able to detect a number of serious errors in dose delivery, leaf bank position and patient thickness during VMAT delivery. Errors in patient setup and single leaf position can only be detected in specific cases.
Asunto(s)
Equipos y Suministros Eléctricos , Errores Médicos , Radiometría/instrumentación , Radioterapia de Intensidad Modulada , Humanos , Neoplasias/radioterapia , Fantasmas de Imagen , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por ComputadorRESUMEN
Electronic portal imaging devices (EPIDs) are frequently used in external beam radiation therapy for dose verification purposes. The aim of this study was to investigate the dose-response characteristics of the EPID in the Unity MR-linac (Elekta AB, Stockholm, Sweden) relevant for dosimetric applications under clinical conditions. EPID images and ionization chamber (IC) measurements were used to study the effects of the magnetic field, the scatter generated in the MR housing reaching the EPID, and inhomogeneous attenuation from the MR housing. Dose linearity and dose rate dependencies were also determined. The magnetic field strength at EPID level did not exceed 10 mT, and dose linearity and dose rate dependencies proved to be comparable to that on a conventional linac. Profiles of fields, delivered with and without the magnetic field, were indistinguishable. The EPID center had an offset of 5.6 cm in the longitudinal direction, compared to the beam central axis, meaning that large fields in this direction will partially fall outside the detector area and not be suitable for verification. Beam attenuation by the MRI scanner and the table is gantry angle dependent, presenting a minimum attenuation of 67% relative to the 90° measurement. Repeatability, observed over two months, was within 0.5% (1 SD). In order to use the EPID for dosimetric applications in the MR-linac, challenges related to the EPID position, scatter from the MR housing, and the inhomogeneous, gantry angle-dependent attenuation of the beam will need to be solved.
Asunto(s)
Imagen por Resonancia Magnética/métodos , Fantasmas de Imagen , Interpretación de Imagen Radiográfica Asistida por Computador/métodos , Radiometría/instrumentación , Planificación de la Radioterapia Asistida por Computador/métodos , Diseño de Equipo , Humanos , Aceleradores de Partículas , Interpretación de Imagen Radiográfica Asistida por Computador/instrumentación , Radiometría/métodos , Dosificación Radioterapéutica , Silicio/químicaRESUMEN
In external beam radiotherapy, electronic portal imaging devices (EPIDs) are frequently used for pre-treatment and for in vivo dose verification. Currently, various MR-guided radiotherapy systems are being developed and clinically implemented. Independent dosimetric verification is highly desirable. For this purpose we adapted our EPID-based dose verification system for use with the MR-Linac combination developed by Elekta in cooperation with UMC Utrecht and Philips. In this study we extended our back-projection method to cope with the presence of an extra attenuating medium between the patient and the EPID. Experiments were performed at a conventional linac, using an aluminum mock-up of the MRI scanner housing between the phantom and the EPID. For a 10 cm square field, the attenuation by the mock-up was 72%, while 16% of the remaining EPID signal resulted from scattered radiation. 58 IMRT fields were delivered to a 20 cm slab phantom with and without the mock-up. EPID reconstructed dose distributions were compared to planned dose distributions using the [Formula: see text]-evaluation method (global, 3%, 3 mm). In our adapted back-projection algorithm the averaged [Formula: see text] was [Formula: see text], while in the conventional it was [Formula: see text]. Dose profiles of several square fields reconstructed with our adapted algorithm showed excellent agreement when compared to TPS.
Asunto(s)
Algoritmos , Imagen por Resonancia Magnética/métodos , Neoplasias/radioterapia , Fantasmas de Imagen , Interpretación de Imagen Radiográfica Asistida por Computador/métodos , Radiometría/métodos , Planificación de la Radioterapia Asistida por Computador/métodos , Humanos , Neoplasias/diagnóstico por imagen , Aceleradores de Partículas , Interpretación de Imagen Radiográfica Asistida por Computador/instrumentación , Cintigrafía , Dosificación Radioterapéutica , Radioterapia de Intensidad Modulada/métodosRESUMEN
At our institution EPID (electronic portal imaging device) dosimetry is routinely applied to perform in vivo dose verification of all patient treatments with curative intent since January 2008. The major impediment of the method has been the amount of work required to produce and inspect the in vivo dosimetry reports (a time-consuming and labor-intensive process). In this paper we present an overview of the actions performed to implement an automated in vivo dosimetry solution clinically. We reimplemented the EPID dosimetry software and modified the acquisition software. Furthermore, we introduced new tools to periodically inspect the record-and-verify database and automatically run the EPID dosimetry software when needed. In 2012, 95% of our 3839 treatments scheduled for in vivo dosimetry were analyzed automatically (27,633 portal images of intensity-modulated radiotherapy (IMRT) fields, 5551 portal image data of VMAT arcs, and 2003 portal images of non-IMRT fields). The in vivo dosimetry verification results are available a few minutes after delivery and alerts are immediately raised when deviations outside tolerance levels are detected. After the clinical introduction of this automated solution, inspection of the detected deviations is the only remaining work. These newly developed tools are a major step forward towards full integration of in vivo EPID dosimetry in radiation oncology practice.
Asunto(s)
Equipos y Suministros Eléctricos , Radiometría/instrumentación , Automatización , Radioterapia de Intensidad ModuladaRESUMEN
The potential for detrimental incidents and the ever increasing complexity of patient treatments emphasize the need for accurate dosimetric verification in radiotherapy. For this reason, all curative treatments are verified, either pretreatment or in vivo, by electronic portal imaging device (EPID) dosimetry in the Radiation Oncology Department of The Netherlands Cancer Institute-Antoni van Leeuwenhoek hospital, Amsterdam, The Netherlands. Since the clinical introduction of the method in January 2005 until August 2009, treatment plans of 4337 patients have been verified. Among these plans, 17 serious errors were detected that led to intervention. Due to their origin, nine of these errors would not have been detected with pretreatment verification. The method is illustrated in detail by the case of a plan transfer error detected in a 5 x 5 Gy intensity-modulated radiotherapy (IMRT) rectum treatment. The EPID reconstructed dose at the isocenter was 6.3% below the planned value. Investigation of the plan transfer chain revealed that due to a network transfer error, the plan was corrupted. 3D analysis of the acquired EPID data revealed serious underdosage of the planning target volume: On average 11.6%, locally up to 20%. This report shows the importance of in vivo (EPID) dosimetry for all treatment plans as well as the ability of the method to assess the dosimetric impact of deviations found.
Asunto(s)
Algoritmos , Radiometría/métodos , Planificación de la Radioterapia Asistida por Computador/métodos , Radioterapia Conformacional/métodos , Tomografía Computarizada por Rayos X/métodos , Pantallas Intensificadoras de Rayos X , Humanos , Dosificación Radioterapéutica , Reproducibilidad de los Resultados , Sensibilidad y EspecificidadRESUMEN
Electronic portal imaging devices (EPIDs) are increasingly used for portal dosimetry applications. In our department, EPIDs are clinically used for two-dimensional (2D) transit dosimetry. Predicted and measured portal dose images are compared to detect dose delivery errors caused for instance by setup errors or organ motion. The aim of this work is to develop a model to predict dose-volume histogram (DVH) changes due to setup errors during breast cancer treatment using 2D transit dosimetry. First, correlations between DVH parameter changes and 2D gamma parameters are investigated for different simulated setup errors, which are described by a binomial logistic regression model. The model calculates the probability that a DVH parameter changes more than a specific tolerance level and uses several gamma evaluation parameters for the planning target volume (PTV) projection in the EPID plane as input. Second, the predictive model is applied to clinically measured portal images. Predicted DVH parameter changes are compared to calculated DVH parameter changes using the measured setup error resulting from a dosimetric registration procedure. Statistical accuracy is investigated by using receiver operating characteristic (ROC) curves and values for the area under the curve (AUC), sensitivity, specificity, positive and negative predictive values. Changes in the mean PTV dose larger than 5%, and changes in V90 and V95 larger than 10% are accurately predicted based on a set of 2D gamma parameters. Most pronounced changes in the three DVH parameters are found for setup errors in the lateral-medial direction. AUC, sensitivity, specificity, and negative predictive values were between 85% and 100% while the positive predictive values were lower but still higher than 54%. Clinical predictive value is decreased due to the occurrence of patient rotations or breast deformations during treatment, but the overall reliability of the predictive model remains high. Based on our predictive model, 2D transit dosimetry measurements can now directly be translated in clinically more relevant DVH parameter changes for the PTV during conventional breast treatment. In this way, the possibility to design decision protocols based on extracted DVH changes is created instead of undertaking elaborate actions such as repeated treatment planning or 3D dose reconstruction for a large group of patients.
Asunto(s)
Algoritmos , Artefactos , Neoplasias de la Mama/radioterapia , Modelos Biológicos , Protección Radiológica/métodos , Radiometría/métodos , Planificación de la Radioterapia Asistida por Computador/métodos , Carga Corporal (Radioterapia) , Simulación por Computador , Humanos , Dosificación Radioterapéutica , Efectividad Biológica Relativa , Reproducibilidad de los Resultados , Sensibilidad y EspecificidadRESUMEN
Electronic portal imaging devices (EPIDs) are not only applied for patient setup verification and detection of organ motion but are also increasingly used for dosimetric verification. The aim of our work is to obtain accurate dose distributions from a commercially available amorphous silicon (a-Si) EPID for transit dosimetry applications. For that purpose, a global calibration model was developed, which includes a correction procedure for ghosting effects, field size dependence and energy dependence of the a-Si EPID response. In addition, the long-term stability and additional buildup material for this type of EPID were determined. Differences in EPID response due to photon energy spectrum changes have been measured for different absorber thicknesses and field sizes, yielding off-axis spectrum correction factors based on transmission measurements. Dose measurements performed with an ionization chamber in a water tank were used as reference data, and the accuracy of the dosimetric calibration model was determined for a large range of treatment conditions. Gamma values using 3% as dose-difference criterion and 3 mm as distance-to-agreement criterion were used for evaluation. The field size dependence of the response could be corrected by a single kernel, fulfilling the gamma evaluation criteria in case of virtual wedges and intensity modulated radiation therapy fields. Differences in energy spectrum response amounted up to 30%-40%, but could be reduced to less than 3% using our correction model. For different treatment fields and (in)homogeneous phantoms, transit dose distributions satisfied in almost all situations the gamma criteria. We have shown that a-Si EPIDs can be accurately calibrated for transit dosimetry purposes.
Asunto(s)
Radiometría/instrumentación , Radiometría/métodos , Silicio , Calibración , Diseño de Equipo , Humanos , Procesamiento de Imagen Asistido por Computador , Modelos Estadísticos , Fantasmas de Imagen , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por Computador , Reproducibilidad de los ResultadosRESUMEN
In 3-dimensional (3D) conformal radiotherapy of parotid gland tumors, little effort is made to avoid the auditory system or the oral cavity. Damage may occur when the ear is located inside the treatment field. The purpose of this study was to design and evaluate an intensity-modulation radiotherapy (IMRT) class solution, and to compare this technique to a 3D conformal approach with respect to hearing loss. Twenty patients with parotid gland cancer were retrospectively planned with 2 different techniques using the original planning target volume (PTV). First, a conventional technique using a wedged beam pair was applied, yielding a dose distribution conformal to the shape of the PTV. Next, an IMRT technique using a fluence map optimization with predefined constraints was designed. A dose of 66 Gy in the PTV was given at the International Commission on Radiation Units and Measures (ICRU) dose prescription point. Dose-volume histograms of the PTV and organs at risk (OARs), such as auditory system, oral cavity, and spinal cord, were compared. The dose in the OARs was lower in the IMRT plans. The mean volume of the middle ear receiving a dose higher than 50 Gy decreased from 66.5% to 33.4%. The mean dose in the oral cavity decreased from 19.4 Gy to 16.6 Gy. The auditory system can be spared if the distance between the inner ear and the PTV is 0.6 cm or larger, and if the overlap between the middle ear and the PTV is smaller than 10%. The maximum dose in the spinal cord was below 40 Gy in all treatment plans. The mean volume of the PTV receiving less than 95% of the prescribed dose increased in the IMRT plan slightly from 3.3% to 4.3 % (p = 0.01). The mean volume receiving more than 107% increased from 0.9% to 2.5% (p = 0.02). It can be concluded that the auditory system, as well as the oral cavity, can be spared with IMRT, but at the cost of a slightly larger dose inhomogeneity in the PTV. The IMRT technique can therefore, in most cases, be recommended as the treatment of choice for the irradiation of parotid tumors.
Asunto(s)
Pérdida Auditiva/etiología , Neoplasias de la Parótida/radioterapia , Radioterapia Conformacional/métodos , Radioterapia de Intensidad Modulada/métodos , Oído/efectos de la radiación , Humanos , Boca/efectos de la radiación , Dosis de Radiación , Planificación de la Radioterapia Asistida por Computador , Radioterapia Conformacional/efectos adversos , Radioterapia de Intensidad Modulada/efectos adversosRESUMEN
Irradiation of the heart is one of the major concerns during radiotherapy of breast cancer. Three-dimensional (3D) treatment planning would therefore be useful but cannot always be performed for left-sided breast treatments, because CT data may not be available. However, even if 3D dose calculations are available and an estimate of the normal tissue damage can be made, uncertainties in patient positioning may significantly influence the heart dose during treatment. Therefore, 3D reconstruction of the actual heart dose during breast cancer treatment using electronic imaging portal device (EPID) dosimetry has been investigated. A previously described method to reconstruct the dose in the patient from treatment portal images at the radiological midsurface was used in combination with a simple geometrical model of the irradiated heart volume to enable calculation of dose-volume histograms (DVHs), to independently verify this aspect of the treatment without using 3D data from a planning CT scan. To investigate the accuracy of our method, the DVHs obtained with full 3D treatment planning system (TPS) calculations and those obtained after resampling the TPS dose in the radiological midsurface were compared for fifteen breast cancer patients for whom CT data were available. In addition, EPID dosimetry as well as 3D dose calculations using our TPS, film dosimetry, and ionization chamber measurements were performed in an anthropomorphic phantom. It was found that the dose reconstructed using EPID dosimetry and the dose calculated with the TPS agreed within 1.5% in the lung/heart region. The dose-volume histograms obtained with EPID dosimetry were used to estimate the normal tissue complication probability (NTCP) for late excess cardiac mortality. Although the accuracy of these NTCP calculations might be limited due to the uncertainty in the NTCP model, in combination with our portal dosimetry approach it allows incorporation of the actual heart dose. For the anthropomorphic phantom, and for fifteen patients for whom CT data were available to test our method, the average difference between the NTCP values obtained with our method and those resulting from the dose distributions calculated with the TPS was 0.1% +/- 0.3% (1 SD). Most NTCP values were 1%-2% lower than those obtained using the method described by Hurkmans et al. [Radiother. Oncol. 62, 163-171 (2002)], using the maximum heart distance determined from a simulator image as a single pre-treatment parameter. A similar difference between the two methods was found for twelve patients using in vivo EPID dosimetry; the average NTCP value obtained with EPID dosimetry was 0.9%, whereas an average NTCP value of 2.2% was derived using the method of Hurkmans et al. The results obtained in this study show that EPID dosimetry is well suited for in vivo verification of the heart dose during breast cancer treatment, and can be used to estimate the NTCP for late excess cardiac mortality. To the best of our knowledge, this is the first study using portal dosimetry to calculate a DVH and NTCP of an organ at risk.
Asunto(s)
Neoplasias de la Mama/radioterapia , Corazón , Imagenología Tridimensional/métodos , Modelos Biológicos , Radiometría/métodos , Planificación de la Radioterapia Asistida por Computador/métodos , Medición de Riesgo/métodos , Algoritmos , Simulación por Computador , Femenino , Humanos , Modelos Estadísticos , Dosificación Radioterapéutica , Valores de Referencia , Factores de RiesgoRESUMEN
The aim of this study was to demonstrate how dosimetry with an amorphous silicon electronic portal imaging device (a-Si EPID) replaced film and ionization chamber measurements for routine pre-treatment dosimetry in our clinic. Furthermore, we described how EPID dosimetry was used to solve a clinical problem. IMRT prostate plans were delivered to a homogeneous slab phantom. EPID transit images were acquired for each segment. A previously developed in-house back-projection algorithm was used to reconstruct the dose distribution in the phantom mid-plane (intersecting the isocenter). Segment dose images were summed to obtain an EPID mid-plane dose image for each field. Fields were compared using profiles and in two dimensions with the y evaluation (criteria: 3%/3 mm). To quantify results, the average gamma (gamma avg), maximum gamma (gamma max), and the percentage of points with gamma < 1(P gamma < 1) were calculated within the 20% isodose line of each field. For 10 patient plans, all fields were measured with EPID and film at gantry set to 0 degrees. The film was located in the phantom coronal mid-plane (10 cm depth), and compared with the back-projected EPID mid-plane absolute dose. EPID and film measurements agreed well for all 50 fields, with (gamma avg) =0.16, (gamma max)=1.00, and (P gamma < 1)= 100%. Based on these results, film measurements were discontinued for verification of prostate IMRT plans. For 20 patient plans, the dose distribution was re-calculated with the phantom CT scan and delivered to the phantom with the original gantry angles. The planned isocenter dose (plan(iso)) was verified with the EPID (EPID(iso)) and an ionization chamber (IC(iso)). The average ratio, (EPID(iso)/IC(iso)), was 1.00 (0.01 SD). Both measurements were systematically lower than planned, with (EPID(iso)/plan(iso)) and (IC(iso)/plan(iso))=0.99 (0.01 SD). EPID mid-plane dose images for each field were also compared with the corresponding plane derived from the three dimensional (3D) dose grid calculated with the phantom CT scan. Comparisons of 100 fields yielded (gamma avg)=0.39, gamma max=2.52, and (P gamma < 1)=98.7%. Seven plans revealed under-dosage in individual fields ranging from 5% to 16%, occurring at small regions of overlapping segments or along the junction of abutting segments (tongue-and-groove side). Test fields were designed to simulate errors and gave similar results. The agreement was improved after adjusting an incorrectly set tongue-and-groove width parameter in the treatment planning system (TPS), reducing (gamma max) from 2.19 to 0.80 for the test field. Mid-plane dose distributions determined with the EPID were consistent with film measurements in a slab phantom for all IMRT fields. Isocenter doses of the total plan measured with an EPID and an ionization chamber also agreed. The EPID can therefore replace these dosimetry devices for field-by-field and isocenter IMRT pre-treatment verification. Systematic errors were detected using EPID dosimetry, resulting in the adjustment of a TPS parameter and alteration of two clinical patient plans. One set of EPID measurements (i.e., one open and transit image acquired for each segment of the plan) is sufficient to check each IMRT plan field-by-field and at the isocenter, making it a useful, efficient, and accurate dosimetric tool.
Asunto(s)
Neoplasias de la Próstata/radioterapia , Radiometría/métodos , Planificación de la Radioterapia Asistida por Computador/métodos , Radioterapia de Intensidad Modulada/métodos , Algoritmos , Calibración , Humanos , Iones , Masculino , Aceleradores de Partículas , Fantasmas de Imagen , Neoplasias de la Próstata/patología , Dosis de Radiación , Dosificación Radioterapéutica , Dispersión de RadiaciónRESUMEN
Many studies have reported dosimetric characteristics of amorphous silicon electronic portal imaging devices (EPIDs). Some studies ascribed a non-linear signal to gain ghosting and image lag. Other reports, however, state the effect is negligible. This study compares the signal-to-monitor unit (MU) ratio for three different brands of EPID systems. The signal was measured for a wide range of monitor units (5-1000), dose-rates, and beam energies. All EPIDs exhibited a relative under-response for beams of few MUs; giving 4 to 10% lower signal-to-MU ratios relative to that of 1000 MUs. This under-response is consistent with ghosting effects due to charge trapping.
Asunto(s)
Procesamiento de Imagen Asistido por Computador/métodos , Radiometría/métodos , Silicio/química , Calibración , Diseño de Equipo , Humanos , Fantasmas de Imagen , Interpretación de Imagen Radiográfica Asistida por Computador , Dosificación Radioterapéutica , Reproducibilidad de los Resultados , Programas Informáticos , Factores de TiempoRESUMEN
Electronic portal imaging devices (EPIDs) can be used to measure a two-dimensional (2D) dose distribution behind a patient, thus allowing dosimetric treatment verification. For this purpose we experimentally assessed the accuracy of a 2D portal dose prediction model based on pencil beam scatter kernels. A straightforward derivation of these pencil beam scatter kernels for portal dose prediction models is presented based on phantom measurements. The model is able to predict the 2D portal dose image (PDI) behind a patient, based on a PDI without the patient in the beam in combination with the radiological thickness of the patient, which requires in addition a PDI with the patient in the beam. To assess the accuracy of portal dose and radiological thickness values obtained with our model, various types of homogeneous as well as inhomogeneous phantoms were irradiated with a 6 MV photon beam. With our model we are able to predict a PDI with an accuracy better than 2% (mean difference) if the radiological thickness of the object in the beam is symmetrically situated around the isocenter. For other situations deviations up to 3% are observed for a homogeneous phantom with a radiological thickness of 17 cm and a 9 cm shift of the midplane-to-detector distance. The model can extract the radiological thickness within 7 mm (maximum difference) of the actual radiological thickness if the object is symmetrically distributed around the isocenter plane. This difference in radiological thickness is related to a primary portal dose difference of 3%. It can be concluded that our model can be used as an easy and accurate tool for the 2D verification of patient treatments by comparing predicted and measured PDIs. The model is also able to extract the primary portal dose with a high accuracy, which can be used as the input for a 3D dose reconstruction method based on back-projection.
Asunto(s)
Algoritmos , Modelos Teóricos , Fantasmas de Imagen , Planificación de la Radioterapia Asistida por Computador , Neoplasias de la Mama/radioterapia , Femenino , Humanos , Dosificación Radioterapéutica , Dispersión de RadiaciónRESUMEN
This study was carried out to determine the stability of the response of amorphous silicon (a-Si)-flat panel imagers for dosimetry applications. Measurements of the imager's response under reference conditions were performed on a regular basis for four detectors of the same manufacturer. We found that the ambient temperature influenced the dark-field, while the gain of the imager signal was unaffected. Therefore, temperature fluctuations were corrected for by applying a "dynamic" darkfield correction. This correction method also removed the influence of a small, irreversible increase of the dark-field current, which was equal to 0.5% of the dynamic range of the imager per year and was probably caused by mild radiation damage to the a-Si array. By applying a dynamic dark-field correction, excellent stability of the response over the entire panel of all imagers of 0.5% (1 SD) was obtained over an observation period up to 23 months. However, two imagers had to be replaced after several months. For one imager, an image segment stopped functioning, while the image quality of the other imager degraded significantly. We conclude that the tested a-Si EPIDs have a very stable response and are therefore well suited for dosimetry. We recommend, however, applying quality assurance tests dedicated to both imaging and dosimetry.
Asunto(s)
Radiometría/instrumentación , Planificación de la Radioterapia Asistida por Computador/instrumentación , Silicio/efectos de la radiación , Calibración , Relación Dosis-Respuesta en la Radiación , Diseño de Equipo , Análisis de Falla de Equipo , Garantía de la Calidad de Atención de Salud/métodos , Garantía de la Calidad de Atención de Salud/normas , Radiometría/normas , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por Computador/métodos , Planificación de la Radioterapia Asistida por Computador/normas , Reproducibilidad de los Resultados , Sensibilidad y Especificidad , TemperaturaRESUMEN
This study was performed to determine the stability of liquid-filled matrix ionization chamber (LiFi-type) electronic portal imaging devices (EPID) for dosimetric purposes. The short- and long-term stability of the response was investigated, as well as the importance of factors influencing the response (e.g., temperature fluctuations, radiation damage, and the performance of the electronic hardware). It was shown that testing the performance of the electronic hardware as well as the short-term stability of the imagers may reveal the cause of a poor long-term stability of the imager response. In addition, the short-term stability was measured to verify the validity of the fitted dose-response curve immediately after beam startup. The long-term stability of these imagers could be considerably improved by correcting for room temperature fluctuations and gradual changes in response due to radiation damage. As a result, the reproducibility was better than 1% (1 SD) over a period of two years. The results of this study were used to formulate recommendations for a quality control program for portal dosimetry. The effect of such a program was assessed by comparing the results of portal dosimetry and in vivo dosimetry using diodes during the treatment of 31 prostate patients. The improvement of the results for portal dosimetry was consistent with the deviations observed with the reproducibility tests in that particular period. After a correction for the variation in response of the imager, the average difference between the measured and prescribed dose during the treatment of prostate patients was -0.7%+/-1.5% (1 SD), and -0.6%+/-1.1% (1 SD) for EPID and diode in vivo dosimetry, respectively. It can be concluded that a high stability of the response can be achieved for this type of EPID by applying a rigorous quality control program.
Asunto(s)
Electrónica Médica/instrumentación , Análisis de Falla de Equipo/métodos , Garantía de la Calidad de Atención de Salud/métodos , Radiometría/instrumentación , Planificación de la Radioterapia Asistida por Computador/instrumentación , Radioterapia/instrumentación , Fluoruros/efectos de la radiación , Compuestos de Litio/efectos de la radiación , Dosis de Radiación , Radiometría/métodos , Radioterapia/métodos , Planificación de la Radioterapia Asistida por Computador/métodos , Reproducibilidad de los Resultados , Sensibilidad y Especificidad , SolucionesRESUMEN
The purpose of this study was to investigate the dose-response characteristics, including ghosting effects, of an amorphous silicon-based electronic portal imaging device (a-Si EPID) under clinical conditions. EPID measurements were performed using one prototype and two commercial a-Si detectors on two linear accelerators: one with 4 and 6 MV and the other with 8 and 18 MV x-ray beams. First, the EPID signal and ionization chamber measurements in a mini-phantom were compared to determine the amount of buildup required for EPID dosimetry. Subsequently, EPID signal characteristics were studied as a function of dose per pulse, pulse repetition frequency (PRF) and total dose, as well as the effects of ghosting. There was an over-response of the EPID signal compared to the ionization chamber of up to 18%, with no additional buildup layer over an air gap range of 10 to 60 cm. The addition of a 2.5 mm thick copper plate sufficiently reduced this over-response to within 1% at clinically relevant patient-detector air gaps (> 40 cm). The response of the EPIDs varied by up to 8% over a large range of dose per pulse values, PRF values and number of monitor units. The EPID response showed an under-response at shorter beam times due to ghosting effects, which depended on the number of exposure frames for a fixed frame acquisition rate. With an appropriate build-up layer and corrections for dose per pulse, PRF and ghosting, the variation in the a-Si EPID response can be reduced to well within +/- 1%.
Asunto(s)
Procesamiento de Imagen Asistido por Computador/métodos , Silicio/química , Aire , Cobre , Relación Dosis-Respuesta en la Radiación , Iones , Aceleradores de Partículas , Fotones , Poliestirenos , Radiometría , Sensibilidad y Especificidad , Factores de TiempoRESUMEN
In this study, we present an algorithm for three-dimensional (3-D) dose reconstruction using portal images obtained with an electronic portal imaging device (EPID). For this purpose an algorithm for 2-D dose reconstruction, which was previously developed in our institution, was adapted. The external contour of the patient was used to correct for absorption of primary photons, but the presence of inhomogeneities was not taken into account. The accuracy of the algorithm was determined by irradiating two anthropomorphic breast phantoms with 6 MV photons. The dose values derived from portal images were compared with results from 3-D dose calculations, which, in turn, were verified with data obtained with an ionization chamber and film dosimetry. It was found that the application of contour information significantly improves the accuracy of 2-D dose reconstruction. If the total dose at the isocenter plane resulting from all treatment beams is reconstructed, the average deviation from the planned dose is 0.1%+/-1.7% (1 SD). If contour information is not available, the differences increase up to +/-20% for the individual beams. In that case, the dose can only be reconstructed with reasonable accuracy when (nearly) opposing beams are used. The average deviation of the 3-D reconstructed dose from the planned dose in the irradiated volume is 1.4%+/-5.4% (1 SD). If the irradiated volume is enclosed by planes less than 5 cm distant from the isocenter plane, then the average deviation is only 0.5%+/-3.4% (1 SD). It can be concluded that the proposed algorithm for a 3-D dose reconstruction allows a determination of the dose at the isocenter plane and the dose-volume histogram with an accuracy acceptable for an independent verification of the treatment.
Asunto(s)
Algoritmos , Neoplasias de la Mama/diagnóstico por imagen , Neoplasias de la Mama/radioterapia , Imagenología Tridimensional/métodos , Intensificación de Imagen Radiográfica/métodos , Radiometría/métodos , Planificación de la Radioterapia Asistida por Computador/métodos , Femenino , Humanos , Imagenología Tridimensional/instrumentación , Fantasmas de Imagen , Protección Radiológica/instrumentación , Protección Radiológica/métodos , Intensificación de Imagen Radiográfica/instrumentación , Radiometría/instrumentación , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por Computador/instrumentación , Reproducibilidad de los Resultados , Medición de Riesgo/métodos , Sensibilidad y EspecificidadRESUMEN
To assess consistency among participants in an European Organisation for Research and Treatment of Cancer (EORTC) phase III trial randomising between irradiation and no irradiation of the internal mammary and medial supraclavicular (IM-MS) lymph nodes, all participating institutes were invited to send data from 3 patients in each arm as soon as they started accrual. The evaluation focused on eligibility, compliance with the radiotherapy guidelines, treatment techniques and dose prescription to the IM-MS region. Nineteen radiotherapy departments provided a total of 111 cases, all being eligible. Minor discrepancies were found in the surgery and pathology data in almost half the patients. Major radiotherapy protocol deviations were very limited: 2 cases of unwarranted irradiation of the supraclavicular region and a significant dose deviation to the internal mammary region in 5 patients. The most frequently observed minor protocol deviation was the absence of delineation of the target volumes in 80% of the patients. By detecting systematic protocol deviations in an early phase of the trial, recommendations made to all the participating institutes should improve the interinstitutional consistency and promote a high-quality treatment.
Asunto(s)
Neoplasias de la Mama/radioterapia , Ensayos Clínicos Controlados Aleatorios como Asunto/normas , Neoplasias de la Mama/patología , Relación Dosis-Respuesta en la Radiación , Femenino , Humanos , Irradiación Linfática/métodos , Calidad de la Atención de Salud , Dosificación Radioterapéutica , Resultado del TratamientoRESUMEN
PURPOSE: With the mean lung dose (MLD) as an estimator for the normal tissue complication probability (NTCP) of the lung, we assessed whether the probability of tumor control of lung tumors might be increased by dose escalation in combination with a reduction of field sizes, thus increasing target dose inhomogeneity while maintaining a constant MLD. METHODS AND MATERIALS: An 8-MV AP-PA irradiation of a lung tumor, located in a cylindrically symmetric lung-equivalent phantom, was modeled using numerical simulation. Movement of the clinical target volume (CTV) due to patient breathing and setup errors was simulated. The probability of tumor control, expressed as the equivalent uniform dose (EUD) of the CTV, was assessed as a function of field size, under the constraint of a constant MLD. The approach was tested for a treatment of a non-small cell lung cancer (NSCLC) patient using the beam directions of the clinically applied treatment plan. RESULTS: In the phantom simulation it was shown that by choosing field sizes that ensured a minimum dose of 95% in the CTV ("conventional" plan) taking into account setup errors and tumor motion, an EUD of the CTV of 43.8 Gy can be obtained for a prescribed dose of 44.2 Gy. By reducing the field size and thus shifting the 95% isodose surface inwards, the EUD increases to a maximum of 68.3 Gy with a minimum dose in the CTV of 55.2 Gy. This increase in EUD is caused by the fact that field size reduction enables escalation of the prescribed dose while maintaining a constant MLD. Further reduction of the field size results in decrease of the EUD because the minimum dose in the CTV becomes so low that it has a predominant effect on the EUD, despite further escalation of the prescribed dose. For the NSCLC patient, the EUD could be increased from an initial 62.2 Gy for the conventional plan, to 83.2 Gy at maximum. In this maximum, the prescribed dose is 88.1 Gy, and the minimum dose in the CTV is 67.4 Gy. In this case, the 95% isodose surface is conformed closely to the "static" CTV during treatment planning. CONCLUSIONS: Iso-NTCP escalation of the probability of tumor control is possible for lung tumors by reducing field sizes and allowing a larger dose inhomogeneity in the CTV. Optimum field sizes can be derived, having the highest EUD and highest minimum dose in the CTV under condition of a constant NTCP of the lungs. We conclude that the concept of homogeneous dose in the target volume is not the best approach to reach the highest probability of tumor control for lung tumors.
Asunto(s)
Neoplasias Pulmonares/radioterapia , Radioterapia Conformacional , Humanos , Fantasmas de Imagen , Probabilidad , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por ComputadorRESUMEN
We compared simulator images of medial tangential fields taken in two positions: (1) with the ipsilateral arm abducted, holding a 'L-bar' armrest and (2) with both arms extended above the head in a forearm support. The average maximum heart distance as well as the central lung distance decreased significantly by 3.4 (SE 0.9) and 4.7 (SE 1.1) mm, respectively, when the new forearm support was used. The estimated normal tissue complication probability for excess cardiac mortality decreased by on average 3.1% (SE 1.3%). For some patients, a greater amount of the axilla was included in the field. We recommend the use of the forearm support during breast cancer treatment with tangential fields to decrease the amount of heart and lung inside the fields.
Asunto(s)
Neoplasias de la Mama/radioterapia , Traumatismos por Radiación/prevención & control , Protección Radiológica , Antebrazo , Corazón , Humanos , Pulmón , Protección Radiológica/instrumentación , Protección Radiológica/métodosRESUMEN
PURPOSE: To determine the intra- and interobserver variation in delineation of the target volume of breast tumors on computed tomography (CT) scans in order to perform conformal radiotherapy. MATERIALS AND METHODS: The clinical target volume (CTV) of the breast was delineated in CT slices by four radiation oncologists on our clinically used delineation system. The palpable glandular breast tissue was marked with a lead wire on 6 patients before CT scanning, whereas 4 patients were scanned without a lead wire. The CTV was drawn by each observer on three separate occasions. Planning target volumes (PTVs) were constructed by expanding the CTV by 7 mm in each direction, except toward the skin. The deviation in the PTV extent from the average extent was quantified in each orthogonal direction for each patient to find a possible directional dependence in the observer variations. In addition, the standard deviation of the intra- and interobserver variation in the PTV volume was quantified. For each patient, the common volumes delineated by all observers and the smallest volume encompassing all PTVs were also calculated. RESULTS: The patient-averaged deviations in PTV extent were larger in the posterior (42 mm), cranial (28 mm), and medial (24 mm) directions than in the anterior (6 mm), caudal (15 mm), and lateral (8 mm) directions. The mean intraobserver variation in volume percentage (5.5%, 1 SD) was much smaller than the interobserver variation (17.5%, 1 SD). The average ratio between the common and encompassing volume for the four observers separately was 0.82, 0.74, 0.82, and 0.80. A much lower combined average ratio of 0.43 was found because of the large interobserver variations. For the observer who placed the lead wire, the intraobserver variation in volume was decreased by a factor of 4 on scans made with a lead wire in comparison to scans made without a lead wire. For the other observers, no improvement was seen. Based on these results, an improved delineation protocol was designed. CONCLUSIONS: Intra- and especially interobserver variation in the delineation of breast target volume on CT scans can be rather large. A detailed delineation protocol making use of CT scans with lead wires placed on the skin around the palpable breast by the delineating observer reduces the intraobserver variation. To reduce the interobserver variation, better imaging techniques and pathology studies relating glandular breast tissue to imaging may be needed to provide more information on the extent of the clinical target volume.