Ann Lab Med ; 41(3): 277-284, 2021 May 01.
Artículo en Inglés | MEDLINE | ID: mdl-33303712

RESUMEN

BACKGROUND: We recently introduced the Barricor (BD, Franklin Lakes, NJ, USA) plasma separation tube, which uses a mechanical separator instead of a gel. We evaluated the effects of using the Barricor tube in a stat (statin) laboratory on the results and turnaround time (TAT) of routine chemical tests. We verified the impact of Barricor tube on reducing TAT and providing results similar to those obtained using serum separator tubes (SSTs). METHODS: We collected venous blood samples from 166 outpatients in Barricor tubes and SSTs and measured 28 routine analytes using an AU5800 instrument (Beckman Coulter, Brea, CA, USA). TAT indexes were compared before and after using Barricor tube. RESULTS: Mean percent differences were <5%, except for alanine aminotransferase , total CO2, high-density lipoprotein, phosphate, total protein, and direct bilirubin. The median TAT decreased from 45 to 38 minutes, and the rate of a TAT >60 minutes decreased from 7.84% to 2.66%, which was approximately one-third of that for SST. The reduction in TAT was attributable to a decrease in centrifugation time. Incomplete clotting and repeated centrifugation, which occurred frequently when using SST, also decreased after using the Barricor tubes. CONCLUSIONS: The Barricor tube is an alternative to SST for routine chemical tests in institutions aiming to reduce TAT, with clinically allowable differences in test results.


Asunto(s)
Pruebas Diagnósticas de Rutina , Alanina Transaminasa , Instituciones de Atención Ambulatoria , Recolección de Muestras de Sangre , Centrifugación , Humanos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA