Artículo en Inglés | MEDLINE | ID: mdl-37566947

RESUMEN

The true coincidence summing correction factor for a Broad Energy Germanium detector has been calculated at far and close geometry set-up using radioactive γ-ray sources. The correction factors were calculated using both experimental and analytical methods. Geant4 simulation was done to calculate the full-energy peak and total efficiencies of the detector. Standard, as well as fabricated mono-energetic γ-ray sources, were used for the γ-ray efficiency measurements. The simulated efficiencies of mono-energetic γ-ray sources were matched to the experimental γ-ray efficiencies by optimizing the detector parameters. The same parameters were used to obtain the full-energy peak and total efficiencies for γ-rays of current interest. Analytical and experimental correction factors were found to agree well with each other. The coincidence summing effect is found to be significant for source-to-detector distances less than 5 cm.