J Hand Surg Br ; 24(6): 654-7, 1999 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-10672797

RESUMEN

A stainless steel external tendon splint was used in repair of cadaver tendons and compared with standard tendon repairs with suture. The splint was combined with a Kessler repair and tested against the Kessler, Becker, and Savage repairs in fresh human cadaver flexor digitorum profundus tendons. Biomechanical testing was done on a tensile testing machine, and load-displacement curves were generated. The repairs using the external tendon splint demonstrated a range of improvement of 32 to 146% in mean maximal tensile strength and a 20 to 185% improvement of mean ultimate tensile strength compared with all other repairs. The external tendon splint is relatively easy to apply to a tendon. The repair is strengthened and becomes capable of withstanding early active range of motion exercises. In vivo testing will be needed to assess the potential clinical usefulness of such a device.


Asunto(s)
Traumatismos de los Dedos/cirugía , Férulas (Fijadores) , Traumatismos de los Tendones/cirugía , Traumatismos de los Dedos/fisiopatología , Humanos , Técnicas In Vitro , Ortopedia/métodos , Acero Inoxidable , Estrés Mecánico , Técnicas de Sutura , Traumatismos de los Tendones/fisiopatología , Tendones/fisiopatología , Resistencia a la Tracción