Quantitative study of signal decay due to magnetic susceptibility interfaces: MRI simulations and experiments.
De Guio, F; Benoit-Cattin, H; Davenel, A.
Afiliación
  • De Guio F; Cemagref, Food Process Engineering Research Unit, CS 64426, 17 avenue de Cucillé, 35044 Rennes, France. francois.de-guio@cemagref.fr
Article en En | MEDLINE | ID: mdl-18002279
ABSTRACT
Characterization of magnetic susceptibility artifacts is required in a wide range of MRI studies. Experiments with a 0.2T magnet and MRI simulations were used to assess the signal decay over echo time at different pixel sizes in the case of air-water interfaces. The specific experimental signal modulation which was reproduced in simulation was interpreted. Concordance between simulation and experiments should make possible further studies of magnetic susceptibility effects with more complex objects as in molecular imaging and in porous medium imaging.
Asunto(s)
Buscar en Google
Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Interpretación de Imagen Asistida por Computador / Aumento de la Imagen / Artefactos / Modelos Biológicos Tipo de estudio: Diagnostic_studies / Prognostic_studies Idioma: En Revista: Annu Int Conf IEEE Eng Med Biol Soc Año: 2007 Tipo del documento: Article
Buscar en Google
Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Interpretación de Imagen Asistida por Computador / Aumento de la Imagen / Artefactos / Modelos Biológicos Tipo de estudio: Diagnostic_studies / Prognostic_studies Idioma: En Revista: Annu Int Conf IEEE Eng Med Biol Soc Año: 2007 Tipo del documento: Article