Your browser doesn't support javascript.
loading
Second international collaborative study evaluating performance characteristics of methods measuring the von Willebrand factor cleaving protease (ADAMTS-13).
Tripodi, A; Peyvandi, F; Chantarangkul, V; Palla, R; Afrasiabi, A; Canciani, M T; Chung, D W; Ferrari, S; Fujimura, Y; Karimi, M; Kokame, K; Kremer Hovinga, J A; Lämmle, B; de Meyer, S F; Plaimauer, B; Vanhoorelbeke, K; Varadi, K; Mannucci, P M.
Afiliación
  • Tripodi A; Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Department of Internal Medicine and Medical Specialties, University and IRCCS Maggiore Hospital, Milan, Italy. armando.tripodi@unimi.it
J Thromb Haemost ; 6(9): 1534-41, 2008 Sep.
Article en En | MEDLINE | ID: mdl-18662260
ABSTRACT

BACKGROUND:

Over the last 4 years ADAMTS-13 measurement underwent dramatic progress with newer and simpler methods.

AIMS:

Blind evaluation of newer methods for their performance characteristics.

DESIGN:

The literature was searched for new methods and the authors invited to join the evaluation. Participants were provided with a set of 60 coded frozen plasmas that were prepared centrally by dilutions of one ADAMTS-13-deficient plasma (arbitrarily set at 0%) into one normal-pooled plasma (set at 100%). There were six different test plasmas ranging from 100% to 0%. Each plasma was tested 'blind' 10 times by each method and results expressed as percentage vs. the local and the common standard provided by the organizer.

RESULTS:

There were eight functional and three antigen assays. Linearity of observed-vs.-expected ADAMTS-13 levels assessed as r2 ranged from 0.931 to 0.998. Between-run reproducibility expressed as the (mean) CV for repeated measurements was below 10% for three methods, 10-15% for five methods and up to 20% for the remaining three. F-values (analysis of variance) calculated to assess the capacity to distinguish between ADAMTS-13 levels (the higher the F-value, the better the capacity) ranged from 3965 to 137. Between-method variability (CV) amounted to 24.8% when calculated vs. the local and to 20.5% when calculated vs. the common standard. Comparative analysis showed that functional assays employing modified von Willebrand factor peptides as substrate for ADAMTS-13 offer the best performance characteristics.

CONCLUSIONS:

New assays for ADAMTS-13 have the potential to make the investigation/management of patients with thrombotic microangiopathies much easier than in the past.
Asunto(s)

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Factor de von Willebrand / Conducta Cooperativa / Proteínas ADAM Tipo de estudio: Clinical_trials Idioma: En Revista: J Thromb Haemost Asunto de la revista: HEMATOLOGIA Año: 2008 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Factor de von Willebrand / Conducta Cooperativa / Proteínas ADAM Tipo de estudio: Clinical_trials Idioma: En Revista: J Thromb Haemost Asunto de la revista: HEMATOLOGIA Año: 2008 Tipo del documento: Article