Midterm stability and hemodynamic performance of a transfemorally implantable nonmetallic, retrievable, and repositionable aortic valve in patients with severe aortic stenosis. Up to 2-year follow-up of the direct-flow medical valve: a pilot study.
Bijuklic, Klaudija; Tuebler, Thilo; Reichenspurner, Hermann; Treede, Hendrik; Wandler, Andreas; Harreld, John H; Low, Reginald I; Schofer, Joachim.
Afiliación
  • Bijuklic K; Medical Care Center Prof Mathey, Prof Schofer, Hamburg University Cardiovascular Center, Hamburg University Heart Center, and Radiologische Allianz, Hamburg, Germany.
Circ Cardiovasc Interv ; 4(6): 595-601, 2011 Dec 01.
Article en En | MEDLINE | ID: mdl-22128202
BACKGROUND: Misplacement during percutaneous aortic valve implantation can be associated with severe complications. The direct flow medical (DFM) valve is repositionable and retrievable; however, the nonmetallic inflatable and conformable design of the valve results in less radial force, which may have an impact on stability and valve function over time. We, therefore, analyzed the midterm stability of the position, shape, and hemodynamic performance of the DFM percutaneous aortic valve. METHODS AND RESULTS: Sixteen symptomatic high-risk for surgery patients with aortic stenosis and a logistic EuroSCORE >20 underwent implantation and were the subject of this analysis. Clinical, echocardiographic, and dual-source multislice computed tomography data were obtained during 2-year follow-up. The 1- and 2-year survival rates were 81% and 69%, respectively. The dual-source multislice computed tomography follow-up indicated no changes in position, diameter, and orifice area of the DFM valve over time. Echocardiography revealed a significant decrease of the mean gradient from baseline (50.1±11.3 mm Hg) to 30 days (19.6±5.7 mm Hg, P<0.001), which remained stable over 2 years. The aortic valve area increased from 0.57±0.15 cm(2) at baseline to 1.47±0.35 cm(2) at 30 days (P<0.001) and did not significantly change during 2-year follow-up. Of the patients, 73% had no aortic regurgitation (AR) and 27% had minimal AR. CONCLUSIONS: In this preliminary series, the 2-year follow-up data of patients, in whom the nonmetallic, repositionable, and retrievable DFM valve was successfully implanted, show stability of the position, shape, and hemodynamic performance, with no AR in most patients.
Asunto(s)

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Válvula Aórtica / Estenosis de la Válvula Aórtica / Índice de Severidad de la Enfermedad / Prótesis Valvulares Cardíacas / Angioplastia / Implantación de Prótesis de Válvulas Cardíacas / Hemodinámica Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Idioma: En Revista: Circ Cardiovasc Interv Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2011 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Válvula Aórtica / Estenosis de la Válvula Aórtica / Índice de Severidad de la Enfermedad / Prótesis Valvulares Cardíacas / Angioplastia / Implantación de Prótesis de Válvulas Cardíacas / Hemodinámica Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Idioma: En Revista: Circ Cardiovasc Interv Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2011 Tipo del documento: Article