Low-dose thallium-201 protocol with a cadmium-zinc-telluride cardiac camera.
Songy, Bernard; Guernou, Mohamed; Lussato, David; Queneau, Mathieu; Geronazzo, Ricardo.
Afiliación
  • Songy B; North Cardiology Hospital, CCN, Saint-Denis, Paris, France. b.songy@ccncardio.com
Nucl Med Commun ; 33(5): 464-9, 2012 May.
Article en En | MEDLINE | ID: mdl-22240935
ABSTRACT

OBJECTIVES:

Thallium-201 is efficient for myocardial perfusion imaging, but leads to relatively high radiation exposure in patients. The purpose of this study was to compare the efficiency of low-dose thallium-201 imaging with cadmium-zinc-telluride (CZT) cameras with regular-dose thallium-201 imaging with conventional cameras.

METHODS:

We prospectively studied 137 consecutive patients referred for stress myocardial perfusion imaging who had previously had a myocardial single-photon emission computed tomography with thallium-201. We injected at stress a low dose of thallium-201 (1.1 MBq/kg, 28 µCi/kg), performed a 5-7 min scan with a CZT camera (GE DNM 530c), and assessed redistribution imaging when the initial images were abnormal. We compared the CZT scan with the conventional dual-head tomographic camera scan taken previously with a regular dose of thallium-201.

RESULTS:

The average delay between both scans was 22 months. The stress dose was 88 ± 16 (2.38 ± 0.43 mCi) versus 125 ± 13 MBq (3.38 ± 0.34 mCi; a 30% reduction). The time for camera acquisition was 6 versus 13 min (a 54% reduction). The myocardial counts were increased two-fold with CZT (mean 446 Kcounts). The quality of CZT images was better in 69% of the cases. There were 59 artifacts with conventional cameras and 29 artifacts with CZT (P<0.01). The diagnostic agreement was calculated in patients without clinical or angiographic changes between both scans (115 patients) and was high (97%). The effective dose at stress was less than 12 mSv. High myocardial counting allowed for further decrease in the injected activity, leading to an effective dose as low as 8 mSv.

CONCLUSION:

With reduced activities of thallium-201 and low effective doses, the CZT camera provides reliable, high-quality imaging.