Your browser doesn't support javascript.
loading
High-performance double-filter soft x-ray diagnostic for measurement of electron temperature structure and dynamics.
McGarry, M B; Franz, P; Den Hartog, D J; Goetz, J A; Thomas, M A; Reyfman, M; Kumar, S T A.
Afiliación
  • McGarry MB; Department of Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706, USA. mbmcgarry@wisc.edu
Rev Sci Instrum ; 83(10): 10E129, 2012 Oct.
Article en En | MEDLINE | ID: mdl-23126950
A new soft x-ray (SXR) T(e) and tomography diagnostic has been developed for MST that can be used for simultaneous SXR spectrum measurement, tomographically reconstructed emissivity, and reconstructed and line-of-sight electron temperature. The diagnostic utilizes high-performance differential transimpedance amplifiers (gain 10(5)-10(9)) to provide fast time response (up to 125 kHz), allowing for the study of plasma structure dynamics. SXR double-foil T(e) measurements are consistent with Thomson scattering. SXR brightness through a variety of filter thicknesses has been combined with charge exchange recombination spectroscopy (CHERS) impurity density measurements to determine the plasma energy spectrum. Magnetic pickup from the fluctuating magnetic fields in the plasma (B̃∼20 gauss at 10-20 kHz) has been dramatically reduced by improving the detector and housing design, so that nanoampere diode currents are now measured without interference from the substantial fluctuating magnetic field incident on the plasma facing surface of the probe.

Texto completo: 1 Base de datos: MEDLINE Tipo de estudio: Diagnostic_studies Idioma: En Revista: Rev Sci Instrum Año: 2012 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Tipo de estudio: Diagnostic_studies Idioma: En Revista: Rev Sci Instrum Año: 2012 Tipo del documento: Article