Your browser doesn't support javascript.
loading
[The external quality assessment schemes for lead in blood organized by the French national agency for medicine and health product safety: a synthesis of 15 years of activity]. / L'évaluation externe de la qualité des analyses de plombémie organisée par l'Agence nationale de sécurité du médicament et des produits de santé : bilan de 15 années de contrôle.
Pineau, Alain; Otz, Jocelyne; Guillard, Olivier; Fauconneau, Bernard; Dumont, Gilles; François-Burg, Elisabeth.
Afiliación
  • Pineau A; Laboratoire de toxicologie, Faculté de pharmacie, Nantes, France.
  • Otz J; Agence nationale de sécurité du médicament et des produits de santé, Saint-Denis, France.
  • Guillard O; Laboratoire de biochimie, Faculté de médecine-pharmacie, Poitiers, France.
  • Fauconneau B; Service de pharmacologie, CHRU, Poitiers, France.
  • Dumont G; Agence nationale de sécurité du médicament et des produits de santé, Saint-Denis, France.
  • François-Burg E; Agence nationale de sécurité du médicament et des produits de santé, Saint-Denis, France.
Ann Biol Clin (Paris) ; 72(1): 49-56, 2014.
Article en Fr | MEDLINE | ID: mdl-24492097
ABSTRACT
In 1992, at the request of the French labor ministry following questions on the ability of medical biology laboratories to satisfactorily measure blood lead level (PbB), a national PbB quality control came into being. Only in 1996 did this external quality control include a number of laboratories sufficient to allow for a significant retrospective evaluation. After fifteen years (1996-2011), The French National Agency for Medicines and Health Products Safety wished to exploit the database collected. The number of participating laboratories went down from 73 to 41. On the other hand, the key finding pertained to the highly improved performance of the laboratories, which was associated with a spread decrease of the results over the entire range of tested PbBs (9 to 700 µg/L). Since 2006, we have observed increasing use of the inductively coupled plasma with mass spectrometry and decreasing use of electrothermal atomic absorption spectrometry. Provided that they rely on identical metrology expertise, the two analytical techniques lead to results on all the tested concentrations that are not statistically different.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Garantía de la Calidad de Atención de Salud / Análisis Químico de la Sangre / Técnicas de Laboratorio Clínico / Seguridad del Paciente / Laboratorios / Plomo / Intoxicación por Plomo Tipo de estudio: Diagnostic_studies / Observational_studies / Risk_factors_studies País/Región como asunto: Europa Idioma: Fr Revista: Ann Biol Clin (Paris) Año: 2014 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Garantía de la Calidad de Atención de Salud / Análisis Químico de la Sangre / Técnicas de Laboratorio Clínico / Seguridad del Paciente / Laboratorios / Plomo / Intoxicación por Plomo Tipo de estudio: Diagnostic_studies / Observational_studies / Risk_factors_studies País/Región como asunto: Europa Idioma: Fr Revista: Ann Biol Clin (Paris) Año: 2014 Tipo del documento: Article