Progress of neutral-beam-aided diagnostics on the HL-2A tokamak.
Rev Sci Instrum
; 85(11): 11E402, 2014 Nov.
Article
en En
| MEDLINE
| ID: mdl-25430309
ABSTRACT
A 32/64-channel Charge eXchange Recombination Spectroscopy (CXRS) and a 7-channel motional Stark effect (MSE) polarimeter have been developed on the HL-2A tokamak. To extract the maximum time resolution of the system, the incidence fibers of the spectrometer are pitch-controlled; and the double-slit fiber bundle can increase the spatial channels with one charge-coupled device detector. The ion temperature and plasma rotation with time and spatial resolutions up to 5 ms and 1 cm are obtained. Sawtooth oscillation, transition from intermediate phase (I phase) to high confinement mode (H mode) can be clearly observed by the CXRS. The spectrometer can be utilized as the main component of the MSE polarimeter, which can effectively overcome the weak Stark effect. The pitch angles of magnetic field are obtained for 7 spatial points covering 24 cm along major radius with time resolution of 40 ms.