Your browser doesn't support javascript.
loading
Evaluation of mobile digital light-emitting diode fluorescence microscopy in Hanoi, Viet Nam.
Chaisson, L H; Reber, C; Phan, H; Switz, N; Nilsson, L M; Myers, F; Nhung, N V; Luu, L; Pham, T; Vu, C; Nguyen, H; Nguyen, A; Dinh, T; Nahid, P; Fletcher, D A; Cattamanchi, A.
Afiliación
  • Chaisson LH; Division of Pulmonary and Critical Care Medicine, San Francisco General Hospital, University of California San Francisco, San Francisco, USA.
  • Reber C; Bioengineering Department, University of California Berkeley, Berkeley, USA.
  • Phan H; Division of Pulmonary and Critical Care Medicine, San Francisco General Hospital, University of California San Francisco, San Francisco, USA.
  • Switz N; Bioengineering Department, University of California Berkeley, Berkeley, USA; Biophysics Graduate Group, University of California Berkeley, Berkeley, California, USA.
  • Nilsson LM; Bioengineering Department, University of California Berkeley, Berkeley, USA.
  • Myers F; Bioengineering Department, University of California Berkeley, Berkeley, USA.
  • Nhung NV; National Lung Hospital, Ba Dinh, Hanoi.
  • Luu L; Hanoi Health Services Department, Ba Dinh, Hanoi.
  • Pham T; Hanoi Lung Hospital, Hai Ba Trung, Hanoi, Viet Nam.
  • Vu C; Hanoi Lung Hospital, Hai Ba Trung, Hanoi, Viet Nam.
  • Nguyen H; Hanoi Lung Hospital, Hai Ba Trung, Hanoi, Viet Nam.
  • Nguyen A; Hanoi Lung Hospital, Hai Ba Trung, Hanoi, Viet Nam.
  • Dinh T; Hanoi Lung Hospital, Hai Ba Trung, Hanoi, Viet Nam.
  • Nahid P; Division of Pulmonary and Critical Care Medicine, San Francisco General Hospital, University of California San Francisco, San Francisco, USA; Curry International Tuberculosis Center, San Francisco General Hospital, University of California San Francisco, San Francisco, California, USA.
  • Fletcher DA; Biophysics Graduate Group, University of California Berkeley, Berkeley, California, USA.
  • Cattamanchi A; Division of Pulmonary and Critical Care Medicine, San Francisco General Hospital, University of California San Francisco, San Francisco, USA; Curry International Tuberculosis Center, San Francisco General Hospital, University of California San Francisco, San Francisco, California, USA.
Int J Tuberc Lung Dis ; 19(9): 1068-72, 2015 Sep.
Article en En | MEDLINE | ID: mdl-26260826
ABSTRACT

SETTING:

Hanoi Lung Hospital, Hanoi, Viet Nam.

OBJECTIVE:

To compare the accuracy of CellScopeTB, a manually operated mobile digital fluorescence microscope, with conventional microscopy techniques.

DESIGN:

Patients referred for sputum smear microscopy to the Hanoi Lung Hospital from May to September 2013 were included. Ziehl-Neelsen (ZN) smear microscopy, conventional light-emitting diode (LED) fluorescence microscopy (FM), CellScopeTB-based LED FM and Xpert(®) MTB/RIF were performed on sputum samples. The sensitivity and specificity of microscopy techniques were determined in reference to Xpert results, and differences were compared using McNemar's paired test of proportions.

RESULTS:

Of 326 patients enrolled, 93 (28.5%) were Xpert-positive for TB. The sensitivity of ZN microscopy, conventional LED FM, and CellScopeTB-based LED FM was respectively 37.6% (95%CI 27.8-48.3), 41.9% (95%CI 31.8-52.6), and 35.5% (95%CI 25.8-46.1). The sensitivity of CellScopeTB was similar to that of conventional LED FM (difference -6.5%, 95%CI -18.2 to 5.3, P = 0.33) and ZN microscopy (difference -2.2%, 95%CI -9.2 to 4.9, P = 0.73). The specificity was >99% for all three techniques.

DISCUSSION:

CellScopeTB performed similarly to conventional microscopy techniques in the hands of experienced TB microscopists. However, the sensitivity of all sputum microscopy techniques was low. Options enabled by digital microscopy, such as automated imaging with real-time computerized analysis, should be explored to increase sensitivity.
Asunto(s)

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Esputo / Tuberculosis Pulmonar / Técnicas Bacteriológicas / Microscopía Fluorescente Tipo de estudio: Diagnostic_studies País/Región como asunto: Asia Idioma: En Revista: Int J Tuberc Lung Dis Año: 2015 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Esputo / Tuberculosis Pulmonar / Técnicas Bacteriológicas / Microscopía Fluorescente Tipo de estudio: Diagnostic_studies País/Región como asunto: Asia Idioma: En Revista: Int J Tuberc Lung Dis Año: 2015 Tipo del documento: Article