Analytical and clinical validation of parathyroid hormone (PTH) measurement in fine-needle aspiration biopsy (FNAB) washings.
Clin Biochem
; 49(1-2): 16-21, 2016 Jan.
Article
en En
| MEDLINE
| ID: mdl-26385495
ABSTRACT
BACKGROUND:
Parathyroid hormone (PTH) quantitation in fine needle aspirate biopsy (FNAB) saline washings complements current modalities for parathyroid tissue localization.OBJECTIVES:
To establish the performance characteristics of the Roche Elecsys intact PTH immunoassay in FNAB needle washings and its diagnostic performance for the identification of parathyroid tissue. DESIGN ANDMETHODS:
Accuracy, precision, reportable range, and analytical specificity and sensitivity for the intact PTH immunoassay in FNAB needle washings were established. For clinical validation, 93 specimens from 79 patients were evaluated. Diagnostic cut-offs were established via receiver operator characteristic (ROC) curve analysis. Performance of PTH in FNAB needle washings was compared to cytology.RESULTS:
Measurement of the PTH in FNAB needle washings demonstrated a matrix interference that was overcome by supplementation of the samples with a protein based matrix prior to analysis. ROC area under the curve (AUC) was 0.96 for PTH in FNAB needle washings. A PTH concentration ≥100pg/mL showed 100% specificity and 82% sensitivity for identifying parathyroid tissue. On histology-confirmed parathyroid specimens, 21/38 (55%) were correctly identified by cytology; whereas 31/38 (82%) were identified by PTH.CONCLUSIONS:
Measurement of PTH in FNAB washings complements cytology for identification of parathyroid tissue. Analytical validation to exclude interference in the PTH immunoassay and proper localization of the parathyroid tissue by ultrasound is necessary to ensure the robustness of the method.Palabras clave