Your browser doesn't support javascript.
loading
A Community Hospital's Evaluation of Alarm Management Safety Factors.
Kurnat-Thoma, Emma; Shah, Kayuri.
Afiliación
  • Kurnat-Thoma E; Author Affiliations: Research Coordinator (Dr Kurnat-Thoma), Research Department, Reston Hospital Center, Reston; and Biomedical Engineering Student (Ms Shah), Department of Biomedical Engineering, Virginia Commonwealth University, Richmond, Virginia.
J Nurs Adm ; 46(12): 675-682, 2016 Dec.
Article en En | MEDLINE | ID: mdl-27851710
ABSTRACT

BACKGROUND:

The Joint Commission's 2014 National Patient Safety Goals required hospitals to evaluate alarm safety in 2014-2015 and implement alarm safety policies.

OBJECTIVE:

The aim of this study was to assess common alarm management safety factors in our 187-bed community hospital.

METHODS:

Two weeks' worth of IV pump report data was evaluated to characterize 33 IV pump alarm types. Hospital and IV pump noise was measured, and an alarm management nurse survey was conducted.

RESULTS:

There were 8731 total IV pump alarms/alerts (24-hour mean, 623.6) across 6 units. The 2-minute idle alarm accounted for 32.4% of all total IV alarms/alerts, suggestive of high levels of nurse multitasking and nurse work interruptions. IV pump volumes contributed to overall hospital noise. Survey data identified patient units and alarm safety practices needing additional support.

CONCLUSIONS:

Characterization of IV pump alarms/alerts is an emerging area of scientific inquiry. Findings indicate the need for organizations to evaluate alarm burden and alarm management safety practices to reduce alarm fatigue risks.
Asunto(s)
Buscar en Google
Base de datos: MEDLINE Asunto principal: Fatiga Auditiva / Administración de la Seguridad / Alarmas Clínicas / Seguridad del Paciente / Ruido / Personal de Enfermería en Hospital Tipo de estudio: Prognostic_studies / Qualitative_research País/Región como asunto: America do norte Idioma: En Revista: J Nurs Adm Año: 2016 Tipo del documento: Article
Buscar en Google
Base de datos: MEDLINE Asunto principal: Fatiga Auditiva / Administración de la Seguridad / Alarmas Clínicas / Seguridad del Paciente / Ruido / Personal de Enfermería en Hospital Tipo de estudio: Prognostic_studies / Qualitative_research País/Región como asunto: America do norte Idioma: En Revista: J Nurs Adm Año: 2016 Tipo del documento: Article