Your browser doesn't support javascript.
loading
In vivo proton observed carbon edited (POCE) 13 C magnetic resonance spectroscopy of the rat brain using a volumetric transmitter and receive-only surface coil on the proton channel.
Kumaragamage, Chathura; Madularu, Dan; Mathieu, Axel P; De Feyter, Henk; Rajah, M Natasha; Near, Jamie.
Afiliación
  • Kumaragamage C; Department of Biomedical Engineering, McGill University, Montreal, QC, Canada.
  • Madularu D; Brain Imaging Centre, Douglas Mental Health University Institute, McGill University, Montreal, QC, Canada.
  • Mathieu AP; Brain Imaging Centre, Douglas Mental Health University Institute, McGill University, Montreal, QC, Canada.
  • De Feyter H; Department of Psychiatry, Faculty of Medicine, McGill University, Montreal, QC, Canada.
  • Rajah MN; Brain Imaging Centre, Douglas Mental Health University Institute, McGill University, Montreal, QC, Canada.
  • Near J; Department of Psychiatry, Faculty of Medicine, McGill University, Montreal, QC, Canada.
Magn Reson Med ; 79(2): 628-635, 2018 02.
Article en En | MEDLINE | ID: mdl-28497464
ABSTRACT

PURPOSE:

In vivo carbon-13 (13 C) MR spectroscopy (MRS) is capable of measuring energy metabolism and neuroenergetics, noninvasively in the brain. Indirect (1 H-[13 C]) MRS provides sensitivity benefits compared with direct 13 C methods, and normally includes a 1 H surface coil for both localization and signal reception. The aim was to develop a coil platform with homogenous B1+ and use short conventional pulses for short echo time proton observed carbon edited (POCE) MRS.

METHODS:

A 1 H-[13 C] MRS coil platform was designed with a volumetric resonator for 1 H transmit, and surface coils for 1 H reception and 13 C transmission. The Rx-only 1 H surface coil nullifies the requirement for a T/R switch before the 1 H preamplifier; the highpass filter and preamplifier can be placed proximal to the coil, thus minimizing sensitivity losses inherent with POCE-MRS systems described in the literature. The coil platform was evaluated with a PRESS-POCE sequence (TE = 12.6 ms) on a rat model.

RESULTS:

The coil provided excellent localization, uniform spin nutation, and sensitivity. 13 C labeling of Glu-H4 and Glx-H3 peaks, and the Glx-H2 peaks were observed approximately 13 and 21 min following the infusion of 1-13 C glucose, respectively.

CONCLUSION:

A convenient and sensitive platform to study energy metabolism and neurotransmitter cycling is presented. Magn Reson Med 79628-635, 2018. © 2017 International Society for Magnetic Resonance in Medicine.