Your browser doesn't support javascript.
loading
Initial experience with the 3.3 Fr Mongoose® pigtail catheter for aortic angiography during patent ductus arteriosus closure in small patients.
Hena, Zachary; Sutton, Nicole J; Gates, Gregory J; Taragin, Benjamin H; Pass, Robert H.
Afiliación
  • Hena Z; Department of Pediatrics, Division of Pediatric Cardiology, The Children's Hospital at Montefiore, Albert Einstein College of Medicine, Bronx, NY, USA.
  • Sutton NJ; Department of Pediatrics, Division of Pediatric Cardiology, The Children's Hospital at Montefiore, Albert Einstein College of Medicine, Bronx, NY, USA.
  • Gates GJ; Department of Pediatrics, Division of Pediatric Cardiology, The Children's Hospital at Montefiore, Albert Einstein College of Medicine, Bronx, NY, USA.
  • Taragin BH; Department of Radiology, Division of Pediatric Cardiology, The Children's Hospital at Montefiore, Albert Einstein College of Medicine, Bronx, NY, USA.
  • Pass RH; Department of Pediatrics, Division of Pediatric Cardiology, The Children's Hospital at Montefiore, Albert Einstein College of Medicine, Bronx, NY, USA.
Ann Pediatr Cardiol ; 10(3): 240-244, 2017.
Article en En | MEDLINE | ID: mdl-28928609
ABSTRACT

BACKGROUND:

Smaller femoral arterial sheaths may be associated with fewer vascular complications. The 3.3 Fr Mongoose® Pediavascular pigtail catheter is a catheter that allows higher flow rates, potentially resulting in improved angiographic quality. We reviewed our experience with this small catheter during patent ductus arteriosus (PDA) closure. MATERIALS AND

METHODS:

Review of patients ≤20 kg in whom the Mongoose® catheter was used during PDA closure from 12/13 to 4/15. Angiographic efficacy and procedural details were compared to ten 4 Fr catheter cases. Comparisons were performed using Mann-Whitney U-test; P < 0.05 was statistically significant.

RESULTS:

Twelve (9 female) patients were catheterized with a 3.3 Fr Mongoose®. Median weight 10.5 kg (range 6.4-18.2), height 81 cm (range 37-111), and body surface area (BSA) 0.47 m2 (range 0.33-0.75) were similar to ten patients (3 females) in the 4 Fr control group (P = NS); median weight 9.9 kg (range 6-16.8), height 80 cm (range 64-102), and BSA 0.46 m2 (range 0.31-0.74). Angiographic quality was subjectively adequate with both with no difference in the median pixel density between the two techniques (3.3 Fr 76.7 [range 33.5-90] and 4 Fr [70; 38-102]; P = NS). Contrast used was similar between the groups (3.3 Fr median 4.2 ml/kg and 4 Fr 4.9 ml/kg; P = NS). Median radiation dose was similar in the two groups (3.3 Fr 28.1 mGy [range 17.2-38] and 4 Fr 38 mGy [range 20.4-58.5]; P = NS). All ducts were closed at latest follow-up (P = NS). No complications were encountered.

CONCLUSIONS:

The 3.3 Fr Mongoose® allowed similar angiography to the 4 Fr pigtail catheter, allowing safe and effective transcatheter PDA closure in small children.
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Idioma: En Revista: Ann Pediatr Cardiol Año: 2017 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Idioma: En Revista: Ann Pediatr Cardiol Año: 2017 Tipo del documento: Article