Automated Processing and Evaluation of Anti-Nuclear Antibody Indirect Immunofluorescence Testing.
Ricchiuti, Vincent; Adams, Joseph; Hardy, Donna J; Katayev, Alexander; Fleming, James K.
Afiliación
  • Ricchiuti V; North Central Division, Laboratory Corporation of America Holdings (LabCorp), Dublin, OH, United States.
  • Adams J; North Central Division, Laboratory Corporation of America Holdings (LabCorp), Dublin, OH, United States.
  • Hardy DJ; North Central Division, Laboratory Corporation of America Holdings (LabCorp), Dublin, OH, United States.
  • Katayev A; Department of Science and Technology, Laboratory Corporation of America Holdings (LabCorp), Elon, NC, United States.
  • Fleming JK; Department of Science and Technology, Laboratory Corporation of America Holdings (LabCorp), Elon, NC, United States.
Front Immunol ; 9: 927, 2018.
Article en En | MEDLINE | ID: mdl-29780386
ABSTRACT
Indirect immunofluorescence (IIF) is considered by the American College of Rheumatology (ACR) and the international consensus on ANA patterns (ICAP) the gold standard for the screening of anti-nuclear antibodies (ANA). As conventional IIF is labor intensive, time-consuming, subjective, and poorly standardized, there have been ongoing efforts to improve the standardization of reagents and to develop automated platforms for assay incubation, microscopy, and evaluation. In this study, the workflow and performance characteristics of a fully automated ANA IIF system (Sprinter XL, EUROPattern Suite, IFA 40 HEp-20-10 cells) were compared to a manual approach using visual microscopy with a filter device for single-well titration and to technologist reading. The Sprinter/EUROPattern system enabled the processing of large daily workload cohorts in less than 8 h and the reduction of labor hands-on time by more than 4 h. Regarding the discrimination of positive from negative samples, the overall agreement of the EUROPattern software with technologist reading was higher (95.6%) than when compared to the current method (89.4%). Moreover, the software was consistent with technologist reading in 80.6-97.5% of patterns and 71.0-93.8% of titers. In conclusion, the Sprinter/EUROPattern system provides substantial labor savings and good concordance with technologist ANA IIF microscopy, thus increasing standardization, laboratory efficiency, and removing subjectivity.