Current state of practice regarding digital radiography exposure indicators and deviation indices: Report of AAPM Imaging Physics Committee Task Group 232.
Dave, Jaydev K; Jones, A Kyle; Fisher, Ryan; Hulme, Katie; Rill, Lynn; Zamora, David; Woodward, Andrew; Brady, Samuel; MacDougall, Robert D; Goldman, Lee; Lang, Susan; Peck, Donald; Apgar, Bruce; Shepard, S Jeff; Uzenoff, Robert; Willis, Charles.
Afiliación
  • Dave JK; Department of Radiology, Thomas Jefferson University, Philadelphia, PA, 19107, USA.
  • Jones AK; Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, TX, 77030, USA.
  • Fisher R; Department of Diagnostic Radiology, The Cleveland Clinic, Beachwood, OH, 44122, USA.
  • Hulme K; Department of Diagnostic Radiology, The Cleveland Clinic, Beachwood, OH, 44122, USA.
  • Rill L; Department of Radiology, University of Florida, Jacksonville Beach, FL, 32250, USA.
  • Zamora D; Department of Radiology, University of Washington, Seattle, WA, 98195, USA.
  • Woodward A; Division of Radiologic Science, University of North Carolina at Chapel Hill, Chapel Hill, NC, 27599, USA.
  • Brady S; Department of Diagnostic Imaging, St. Jude Children's Research Hospital, Memphis, TN, 38105, USA.
  • MacDougall RD; Department of Radiology, Boston Children's Hospital, Boston, MA, 02115, USA.
  • Goldman L; Department of Radiology, Hartford Hospital, Hartford, CT, 06102, USA.
  • Lang S; Department of Radiology, Henry Ford Health System, Detroit, MI, 48202, USA.
  • Peck D; Department of Radiology, Henry Ford Health System, Detroit, MI, 48202, USA.
  • Apgar B; Agfa Healthcare, Greenville, SC, 29601, USA.
  • Shepard SJ; Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, TX, 77030, USA.
  • Uzenoff R; Fujifilm Medical Systems U.S.A., Inc., Stamford, CT, 06902, USA.
  • Willis C; Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, TX, 77030, USA.
Med Phys ; 45(11): e1146-e1160, 2018 Nov.
Article en En | MEDLINE | ID: mdl-30255505
ABSTRACT
Beginning with the advent of digital radiography systems in 1981, manufacturers of these systems provided indicators of detector exposure. These indicators were manufacturer-specific, and users in facilities with equipment from multiple manufacturers found it a challenge to monitor and manage variations in indicated exposure in routine clinical use. In 2008, a common definition of exposure index (EI) was realized in International Electrotechnical Commission (IEC) International Standard 62494-1 Ed. 1, which also introduced and defined the deviation index (DI), a number quantifying the difference between the detector EI for a given radiograph and the target exposure index (EIT ). An exposure index that differed by a constant from that established by the IEC and the concept of the deviation index also appear in American Association of Physicists in Medicine (AAPM) Report No. 116 published in 2009. The AAPM Report No. 116 went beyond the IEC standard in supplying a table (Table II in the report of TG-116) titled "Exposure Indicator DI Control Limits for Clinical Images," which listed suggested DI ranges and actions to be considered for each range. As the IEC EI was implemented and clinical DI data were gathered, concerns were voiced that the DI control limits published in the report of TG-116 were too strict and did not accurately reflect clinical practice. The charge of task group 232 (TG-232) and the objective of this final report was to investigate the current state of the practice for CR/DR Exposure and Deviation Indices based on AAPM TG 116 and IEC-62494, for the purpose of establishing achievable goals (reference levels) and action levels in digital radiography. Data corresponding to EI and DI were collected from a range of practice settings for a number of body parts and views (adults and pediatric radiographs) and analyzed in aggregate and separately. A subset of radiographs was also evaluated by radiologists based on criteria adapted from the European Guidelines on Quality Criteria for Diagnostic Radiographic Images from the European Commission. Analysis revealed that typical DI distribution was characterized by a standard deviation (SD) of 1.3-3.6 with mean DI values substantially different from 0.0, and less than 50% of DI values fell within the significant action limits proposed by AAPM TG-116 (-1.0 ≤ DI ≤ 1.0). Recommendations stemming from this analysis include targeting a mean DI value of 0.0 and action limits at ±1 and ±2 SD of the DI based on actual DI data of an individual site. EIT values, DI values, and associated action limits should be reviewed on an ongoing basis and optimization of DI values should be a process of continuous quality improvement with a goal of reducing practice variation.