Interchangeability of corneal curvature and asphericity measurements provided by three different devices.
Piñero, David P; Soto-Negro, Roberto; Ruiz-Fortes, Pedro; Pérez-Cambrodí, Rafael J; Fukumitsu, Hideki.
Afiliación
  • Piñero DP; Department of Optics, Pharmacology and Anatomy, University of Alicante, San Vicente del Raspeig 03690, Alicante, Spain.
  • Soto-Negro R; Department of Ophthalmology (OFTALMAR), Vithas Medimar International Hospital, Alicante 03014, Spain.
  • Ruiz-Fortes P; Department of Ophthalmology (OFTALMAR), Vithas Medimar International Hospital, Alicante 03014, Spain.
  • Pérez-Cambrodí RJ; Department of Ophthalmology (OFTALMAR), Vithas Medimar International Hospital, Alicante 03014, Spain.
  • Fukumitsu H; Department of Ophthalmology (OFTALMAR), Vithas Medimar International Hospital, Alicante 03014, Spain.
Int J Ophthalmol ; 12(3): 412-416, 2019.
Article en En | MEDLINE | ID: mdl-30918809
AIM: To evaluate the interchangeability of keratometric and asphericity measurements provided by three measurement systems based on different optical principles. METHODS: A total of 40 eyes of 40 patients with a mean age of 34.1y were included. In all cases, a corneal curvature analysis was performed with IOL-Master (IOLM), iDesign 2 (ID2), and Sirius systems (SIR). Differences between instruments for flattest (K1) and steepest (K2) keratometric readings, as well as for magnitude and axis of corneal astigmatism were analyzed. Likewise, differences in asphericity (Q) between SIR and ID2 were also evaluated. RESULTS: Mean differences between devices for K1 were 0.20±0.21 (P<0.001), -0.12±0.36 (P=0.046) and -0.32±0.36 D (P<0.001) for the comparisons IOLM-SIR, IOLM-ID2 and SIR-ID2, respectively. The ranges of agreement for these comparisons between instruments were 0.41, 0.70, and 0.70 D. For K2, mean differences were 0.31±0.33 (P<0.001), -0.08±0.43 (P=0.265) and -0.39±0.38 D (P<0.001), with ranges of agreement of 0.65, 0.84, and 0.74 D. Concerning magnitude of astigmatism, ranges of agreement were in the limit of clinical relevance (0.49 D, P=0.011; 0.55 D, P=0.386; 0.43 D, P=0.05). In contrast, ranges of agreement were clinically relevant for astigmatic axis (26.68°, 33.83° and 18.37°, P≥0.121) and for Q between SIR and ID2 (0.16, P<0.001). CONCLUSION: The keratometric corneal power, astigmatic axis and asphericity measurements provide by the three systems evaluated cannot be considered as interchangeable, whereas measurements of corneal astigmatism obtained with SIR and ID2 can be considered as interchangeable for clinical purposes.