Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems.
Richter, Anne; Wegener, Sonja; Breuer, Kathrin; Razinskas, Gary; Weick, Stefan; Exner, Florian; Bratengeier, Klaus; Flentje, Michael; Sauer, Otto; Polat, Bülent.
Afiliación
  • Richter A; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany. richter_a3@ukw.de.
  • Wegener S; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
  • Breuer K; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
  • Razinskas G; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
  • Weick S; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
  • Exner F; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
  • Bratengeier K; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
  • Flentje M; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
  • Sauer O; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
  • Polat B; Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
Radiat Oncol ; 16(1): 213, 2021 Nov 06.
Article en En | MEDLINE | ID: mdl-34742291
BACKGROUND: To implement a tangential treatment technique for whole breast irradiation using the Varian Halcyon and to compare it with Elekta Synergy Agility plans. METHODS: For 20 patients two comparable treatment plans with respect to dose coverage and normal tissue sparing were generated. Tangential field-in-field treatment plans (Pinnacle/Synergy) were replanned using the sliding window technique (Eclipse/Halcyon). Plan specific QA was performed using the portal Dosimetry and the ArcCHECK phantom. Imaging and treatment dose were evaluated for treatment delivery on both systems using a modified CIRS Phantom. RESULTS: The mean number of monitor units for a fraction dose of 2.67 Gy was 515 MUs and 260 MUs for Halcyon and Synergy Agility plans, respectively. The homogeneity index and dose coverage were similar for both treatment units. The plan specific QA showed good agreement between measured and calculated plans. All Halcyon plans passed portal dosimetry QA (3%/2 mm) with 100% points passing and ArcCheck QA (3%/2 mm) with 99.5%. Measurement of the cumulated treatment and imaging dose with the CIRS phantom resulted in lower dose to the contralateral breast for the Halcyon plans. CONCLUSIONS: For the Varian Halcyon a plan quality similar to the Elekta Synergy device was achieved. For the Halcyon plans the dose contribution from the treatment fields to the contralateral breast was even lower due to less interleaf transmission of the Halcyon MLC and a lower contribution of scattered dose from the collimator system.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Aceleradores de Partículas / Garantía de la Calidad de Atención de Salud / Planificación de la Radioterapia Asistida por Computador / Fantasmas de Imagen / Radioterapia de Intensidad Modulada / Neoplasias Tipo de estudio: Etiology_studies / Observational_studies / Risk_factors_studies Idioma: En Revista: Radiat Oncol Asunto de la revista: NEOPLASIAS / RADIOTERAPIA Año: 2021 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Aceleradores de Partículas / Garantía de la Calidad de Atención de Salud / Planificación de la Radioterapia Asistida por Computador / Fantasmas de Imagen / Radioterapia de Intensidad Modulada / Neoplasias Tipo de estudio: Etiology_studies / Observational_studies / Risk_factors_studies Idioma: En Revista: Radiat Oncol Asunto de la revista: NEOPLASIAS / RADIOTERAPIA Año: 2021 Tipo del documento: Article