Stability of implant-abutment connection in three different systems after fatigue test.
J Indian Prosthodont Soc
; 22(4): 338-342, 2022.
Article
en En
| MEDLINE
| ID: mdl-36511067
Aim: Abutment screw loosening of implant-supported prosthesis causes a mismatch between the abutment and the implant. This screw loosening is influenced by the implant-abutment connection type, however, with contradictory results reported in different studies. The present study evaluates the stability of abutment-implant connections in three different systems before and after the fatigue test. Settings and Design: Thirty implants (4.3 mm in diameter and 12 mm in length) were divided into three groups of 10: Implantium, Zimmer, and straight internal hexagonal connection (SIC) implants. Materials and Methods: Two torques of 35 Ncm with an interval of 10 min were applied, followed by measuring removal torque value (RTV). The samples were re-torqued and then underwent a simulation of 1-year chewing clinical performance of dental implant under axial force of 400 N, with a frequency of 8 Hz (one million cycles). After fatigue test, the RTV was calculated and recorded. Statistical Analysis: The mean RTVs obtained before and after cyclic load were analyzed by SPSS version 22 software using multivariate analysis. Results: Significant differences in RTV and role of cyclic loading were found between SIC and Implantium groups (P = 0.006 and 0.021, respectively), as well as between Zimmer and SIC groups (P = 0.032 and 0.006, respectively), but not between Zimmer and Implantium groups (P = 0.771 and 0.248, respectively). Conclusion: The type of connection could affect the screw loosening, the preload loss, and the implant component stability. SIC group revealed the highest RTVs before and after cyclic loading.
Palabras clave
Texto completo:
1
Base de datos:
MEDLINE
Asunto principal:
Pilares Dentales
/
Diseño de Implante Dental-Pilar
Idioma:
En
Revista:
J Indian Prosthodont Soc
Año:
2022
Tipo del documento:
Article