Linear relationship between absorbed radiation dose and pulse rate during fluoroscopy.
Yecies, Todd S; Chun, Brian; Semins, Michelle J; Chaudhry, Rajeev.
Afiliación
  • Yecies TS; Department of Urology, Palo Alto Foundation Medical Group, Palo Alto, CA, USA.
  • Chun B; Department of Urology, UPMC Children's Hospital of Pittsburgh, 3471 Fifth Avenue. Suite 700, Pittsburgh, PA, 15213, USA. chunb@upmc.edu.
  • Semins MJ; Department of Urology, West Virginia University School of Medicine, Morgantown, WV, USA.
  • Chaudhry R; Department of Urology, UPMC Children's Hospital of Pittsburgh, 3471 Fifth Avenue. Suite 700, Pittsburgh, PA, 15213, USA.
World J Urol ; 41(1): 269-274, 2023 Jan.
Article en En | MEDLINE | ID: mdl-36525105
ABSTRACT

PURPOSE:

To identify the relationship between fluoroscopy pulse rate and absorbed radiation dose. We compared absorbed radiation dose with common proxy measurements such as fluoroscopy time and C-arm reported dose.

METHODS:

Using a simulated patient model, 60 s fluoroscopy exposures were performed using pulse rates of 30, 8, 4, 2, and 1 pulse(s) per second. Each experiment was performed with both standard and low-dose settings using a GE OEC 9800 plus C-arm. Landauer nanoDot™ OSL dosimeters were used to measure the absorbed radiation dose.

RESULTS:

Fluoroscopy pulse rate and absorbed radiation dose demonstrated a linear correlation for both standard (R2 = 0.995, p < 0.001) and low-dose (R2 = 0.998, p < 0.001) settings. For any given pulse rate, using the low-dose setting reduced absorbed radiation dose by 58 ± 2.8%. Fluoroscopy time demonstrated a linear relationship with absorbed radiation dose for both standard (R2 = 0.996, p < 0.001) and low-dose (R2 = 0.991, p < 0.001) settings, but did not change with use of the low-dose setting. C-arm reported radiation dose correlated linearly with absorbed dose (R2 = 0.999) but consistently under-estimated measured values by an average of 49 ± 3.5%. Using a combination of 1 pulse-per-second and low-dose fluoroscopy, absorbed dose was reduced by 97.7 ± 0.1% compared to standard dose and 30 pulse-per-second settings.

CONCLUSION:

Absorbed radiation dose decreases linearly with fluoroscopy pulse rate during equivalent exposure times. Adjusting fluoroscopy pulse rate and utilizing low-dose settings significantly reduces overall absorbed radiation exposure by up to 98%.