Rationale for and use of the lumenless 3830 pacing lead.
Richardson, Travis D; Himes, Adam; Marshall, Mark; Crossley, George H.
Afiliación
  • Richardson TD; Division of Cardiovascular Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Himes A; Medtronic, Inc., Minneapolis, Minnesota, USA.
  • Marshall M; Medtronic, Inc., Minneapolis, Minnesota, USA.
  • Crossley GH; Division of Cardiovascular Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
J Cardiovasc Electrophysiol ; 34(3): 769-774, 2023 03.
Article en En | MEDLINE | ID: mdl-36682066
ABSTRACT
Most currently available pacing and defibrillation leads utilize a stylet-based design that facilitates implantation. This has advantages, but also increases the lead diameter and adds the potential for metal fatigued-based conductor failure. A systematic literature search was conducted, and the authors add their twenty-year experience with this lead design. The global experience with lumenless leads was reviewed both for "standard" positioning and with conduction system pacing. Methods for both placement and system modification are reviewed. Lumenless leads have the potential to improve the durability of endocardial pacing and facilitate conduction system pacing.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Marcapaso Artificial Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2023 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Marcapaso Artificial Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2023 Tipo del documento: Article